Understanding Test Results and Messages
About Quality Control Testing
What is a Quality Control Test ?
To ensure proper monitoring function, it is necessary to regularly perform a quality control test.
Use one of the 
 Control Solutions GC700 when testing your 
 Blood Glucose 
Monitoring System GM782 in the Control Solution Mode. If the test result is within the Control 
Solution Range printed on the strip vial label, the 
Blood Glucose Monitoring System 
GM782 passes the Quality Control Test and your 
 Blood Glucose Monitoring System 
GM782 is working properly.
TM
TM
Rightest
Rightest
TM
Rightest
TM
Rightest
 
Control Solution Range:
Example of Control Solution Range 
printed on your test strip vial label.
Control Solution Range mg/dL
Low:                 28-48
Normal:          107-145
High:              315-426  
NOTE
- Normal of 
 Control Solution GC700 is provided in your starter kit. If you want to 
purchase, High or another Normal control solutions, please contact customer service.
TM
 Rightest
The 
 Meter
displays results between 10 
and 600 mg/dL.
If the test result is below 10 mg/dL , “      ” will appear on the screen. Please 
repeat your test with by a new test strip.
If you still get a “      ” result, contact your healthcare professional.
If the test result is above 600 mg/dL, “      ” will appear on the screen. Please 
repeat your test again with a new test strip. 
If you still get a “      ” result, contact your healthcare professional.
TM
 Rightest  Blood Glucose
 GM782