Bionime Corporation GM782 Manuale Utente
- Avoid contact with dripping or splashing liquids.
- The minimum blood sample size to test using the
- The minimum blood sample size to test using the
Blood Glucose Monitoring System
is 0.75µL: ( )
TM
Rightest
GM782
5
4
Caution
Caution
Blood sample size above 3.0µL might contaminate the test strip port and the meter.
Blood sample size below 0.75µL may cause an inaccurate result or may prevent a meter reading.
An Er4 reading will be displayed if the sample size is too small. In this case, repeat the test with a
new test strip.
Blood sample size below 0.75µL may cause an inaccurate result or may prevent a meter reading.
An Er4 reading will be displayed if the sample size is too small. In this case, repeat the test with a
new test strip.
Sample Size Example
Important Safety Notes:
- Users should wash their hands thoroughly with soap and water before and after handling
- Users should wash their hands thoroughly with soap and water before and after handling
the meter, lancing device, or test strips.
0.75µL 1.0µL 1.5µL 2.0µL 3.0µL
- Before using the
Blood Glucose Monitoring System
to test your blood glucose,
please read all of the instructions and conduct all of the tests including the quality control test
- Please perform the quality control test regularly to make sure the test results are accurate.
TM
TM
Rightest
GM782
Rightest
GS700 Other brands’ test strips should not be used under any circumstances. The
use of other brands’ strips may give inaccurate results.
TM
The Rightest Blood Glucose Monitoring System GM782 is intended for in vitro diagnostic use
TM
only. The Rightest Blood Glucose test results using fresh capillary whole blood samples from
the fingertip, palm and forearm are calibrated to be the equivalent to that of plasma testing.
Venous, neonatal or arterial blood may also be used only if drawn by healthcare professionals.
the fingertip, palm and forearm are calibrated to be the equivalent to that of plasma testing.
Venous, neonatal or arterial blood may also be used only if drawn by healthcare professionals.
TM
Rightest
GM782
TM
- If the Rightest Blood Glucose Meter GM782 and Test Strips GS700 are exposed to a substantial
change in temperature, please wait 30 minutes before measurement.
- Follow all environmental protection regulations when disposing of batteries.
TM
Rightest
GM782
(Refer to page 37).
- The
Blood Glucose Meter
can only be used with
Blood Glucose
Test Strips
.
-
- The
Blood Glucose Monitoring System
is intended for self-testing. It should not
be used to diagnose diabetes mellitus.
The system has been tested with neonatal blood. As a matter of good clinical practice, caution is
advised in the interpretation of neonate glucose value below 50 mg/dL. Please follow the
recommendations for follow-up care that have been set by your intuition for critical glucose values
in neonates. Results for neonates exhibiting symptoms of galactosemia should be confirmed by
laboratory tests.
advised in the interpretation of neonate glucose value below 50 mg/dL. Please follow the
recommendations for follow-up care that have been set by your intuition for critical glucose values
in neonates. Results for neonates exhibiting symptoms of galactosemia should be confirmed by
laboratory tests.