Bionime Corporation GM782 Manuale Utente

Pagina di 32
15
14
The 
 Blood Glucose Meter G 782 can only be used with 
 Blood Glucose Test 
Strips GS700 and 
 Control Solution GC700. The use of other test strips or control 
solutions can lead to incorrect results.
TM
TM
Rightest
M
Rightest
TM
Rightest
PRECAUTION
- Close the 
  lood Glucose Test Strip GS700 vial immediately after removing a 
test strip.
- Do not reuse 
.
- Do not use expired 
 Discard the vial of test strips after 4 months from opening.
- Store the 
 in a cool ( 39-86°F or 4-30°C ) and 
dry location (< 90% relative humidity). Do not expose to direct sunlight or heat.
TM
Rightest B
TM
Rightest  Blood Glucose Test Strip GS700
TM
Rightest  Blood Glucose Test Strip GS700
TM
When you open a new vial of Rightest  Blood Glucose Test Strip GS700, record the date 
on the vial.
TM
Rightest  Blood Glucose Test Strip GS700
TM
- If the Rightest Blood Glucose Meter 
 and Test Strips are exposed to a substantial 
change in temperature, please wait 30 minutes before measurement.
TM
- For detailed information, please refer to the Rightest  Blood Glucose Test Strip GS700 
Package Insert.
 
GM782
The 
 Blood Glucose Test Strip GS700
TM
Rightest
View Window
This window is yellow 
before applying blood 
sample. It gradually turns 
red when filled with blood.
Hand Bar
Grip here to insert 
test strip into the 
strip port of auto 
code key.
Sample Entry
µL
Apply a drop of blood or control solution here. 
The test request is only 0.75 
 of blood. 
Noble Metal Electrodes
Electrochemical sensor
Indication Symbol
Insert strip with indication symbol 
up and toward meter.
The 
 Blood Glucose Test Strip GS700
TM
Rightest
Electrode Contacts
Sensing signal output terminals.