Technical Information: 
Temperature Probes (RTD) Specification 
Parameter  
Specification  
Sensor Type  
1000 Ohm RTD Temperature Sensor 
Sensor Curve Type  
TC=3850 ppm/K 
Sensor Self-Heating  
0.4K/mW at 0 C  
Sensor Accuracy  
+/- 0.15 C minimum including cable 
and connector  
Range  
10 to 45 C(50 to 113 F)  
Resolution  
0.01 C (0.018 F)  
Accuracy  
+/- 0.1 C (0.18 F)  
Host 
Sampling 
Rate 
1 Hz to 40Hz in integer value 
increments 
Wire 
Type 
24-28 AWG twisted pairs with 
overall shield  
Max Sensor Cable Length 
6 Feet  
VENDYS II Machine Specification 
Parameter Specification 
Description 
 
Min Inflated Pressure  
10 mmHg  
Minimum  pressure  to 
which a cuff can be 
inflated and held  
Max Inflated Pressure  
270 mmHg  
Maximum  pressure  to 
which a cuff can be 
inflated and held  
Pressure Resolution  
1 mmHg  
Pressure  reported  to 
host will be to nearest 1