eResearchTechnology GmbH AM3G01 Manuale Utente

Pagina di 42
Page 33/42
Peak Flow Meter AM3 - Instructions for Use for Trial xxx
Version 00.03 • Date 04NOV2013
General Safety Precautions 
Guidance and Manufacturer‘s Declaration –  Electromagnetic Immunity
The AM3 GSM is intended for use in the electromagnetic environment specified below. The customer or the 
user of the AM3 GSM should ensure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance 
level
Electromagnetic environment - guidance
Portable and mobile RF communications 
equipment should be used no closer to any 
part of the AM3 GSM, including cables, 
that the recommended separation distance 
calculated from the equation applicable to 
the frequency of the transmitter:
Recommended protection distance:
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
3 Vrms
d = 1.17 1/V * √P
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2,5 GHz
3 V/m
d = 1.17 m/V * √P  for 80 MHz to 800 MHz
d = 2,33 m/V * √P  for 800 MHz to 2.5 GHz
where P is the maximum output power 
rating of the transmitter in watts (W) acc. 
to the transmitter manufacturer and d is 
the recommended separation distance in 
meters (m). 
Field strengths from fixed RF transmitters, 
as determined by an electromagnetic site 
survey,
a
 should be less than the compli-
ance level in each frequency range.
b
 
Interference may occur in the vicinity 
of equipment marked with the following 
symbol:
                 
NOTE 1          At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2          These guidelines may not apply in all situations. Electromagnetic propagation is affected by 
absorption and reflections from structures, objects and people.
a         Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones 
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted 
theroretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an 
electromagnetic site survey should be considered. If the measured field strength in the location in which the 
AM3 GSM is used exceeds the applicable RF compliance level above, the AM3 GSM should be observed to 
verify normal operation. If abnormal performance is observed, additional measures may be necessary, such 
as reorienting or relocating the AM3 GSM.
b         Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.