Truly Instrument Ltd. DW702 Manuale Utente

Pagina di 1
SIZE:72*97MM  正面
SIZE:72*97MM  反面
21
“Er P” displayed
Fail to inflate
Check if the cuff is properly 
connected.Replace the air 
tube if it is broken. 
“Er 1” displayed
Deflate too rapidly Return for servicing.
“Er 2” displayed
Movement during 
measurement
Do not move during 
measurement
Signal interference Remove interfering source 
e.g. mobile phones, 
magnets.
“Er 3” displayed
Incorrect results
Measure again.
Note: If your problem cannot be solved by the above, 
           consult your store of purchase. Do not disassemble 
           the unit.
Problem                 Cause 
     Solution
No display when 
you press the Power
/Start button
Have the batteries 
run out?
Are the battery 
polarities wrong?
Replace new batteries.
Re-align batteries in correct 
polarities.
TROUBLESHOOTING
15
3.  Remove tight fitting clothing from 
     you upper arm and thick clothing. 
     Do not roll up your sleeve if it is 
     too tight.
4.  Place your arm on a table so that the 
     cuff will be at the same level as your 
     heart.
TAKING  MEASUREMENT
Note:
   -    Avoid smoking, eating or exercising for 30 minutes before 
        taking measurement.
   -    Do not move or talk during measurement.
   -    Measurement can be taken on either arm.
3.1  Sitting Correctly
1.  Relax
2.  Sit upright in a chair with your both feet on the floor.
3.2  Applying the Wrist Cuff
1.   Place your elbow on a table and palm facing upward, and 
      wrap the arm cuff around the upper left arm.  
2.   Pull on the end of the cuff until it wraps securely around 
      your upper wrist.  Do not over tighten the cuff. Allow 1-2 
      cm or ½ inch between the bottom of the cuff and your 
      wrist joint.
3.3  User Setting
3.   Make sure the cuff is at the same level as your heart.
Press ”PAGE   ”  button to switch between User Page 
1 ,2,3 and 4.
This unit allows a maximum of 4 users to store 
measurement records in different User Pages.
16
1.  Press the “     ” button. The cuff will automatically inflate 
     and the measurement will start.
2.  When the measurement is complete, the cuff will 
     automatically deflate and your systolic and diastolic 
     pressure values and pulse rate will display. 
3.  Remove the wrist
 
cuff and take note of your measurement 
     results.
3.4  Taking Measurement
Note: You can press the “     ” button to stop measurement at 
           any time.
19
18
17
20
3.5  Heartbeat Readings
Besides Systolic and Diastolic values, 
Pulse will also be displayed after a 
measurement
Pulse
Irregular Heartbeat
This unit has a unique feature that alerts user of irregular 
heartbeat detection during measurement.
Note: An irregular heartbeat is defined as a 
           heartbeat rhythm that has a variation 
           of more than 25% from the average 
           rhythm detected during 
           measurement.
If such irregular rhythm occurs twice or 
more during the same measurement, the 
display screen will show the pulse value 
and “IH” symbol alternately.Example:  
(90 / IH)
Note:  Please consult your doctor if 
            irregular heartbeat occurs often.
Regular but Below Normal Heartbeat
If the monitor detects a pulse rate below 
60 beats per minute during measurement, 
the monitor will display the pulse value and 
“LO” symbol alternately. Example:  (58/LO)
If the monitor detects a pulse rate above 
100 beats per minute during measurement, 
the monitor will display the pulse value and
 “HI” symbol alternately. Example:  (102 / HI)
Regular but Above Normal Heartbeat
1.  Press the “PAGE   ” button to select the User Page you want 
     to delete.
4.3  Delete Past Readings
User can delete all the past readings stored in each User 
Page.
2.  Press and hold the “   MEMO” & “PAGE    ” button 
     simultaneously for 5 seconds. Symbol of “EE EE” will be 
     displayed and all past readings stored in that User Page 
     will be deleted.
 
1.  Press the “   MEMO”  button to enter memory function.
2.  If the User Page is not what you want to view, press the 
     “PAGE    ”  button to select.
After measurement, the last set of measured values will be 
saved automatically after measurement when the unit is 
turned off by pressing “     ” button or by automatic power-off 
after 3 minutes.
4.1  View Past Readings
MEMORY  FUNCTIONS
3.  Press the “   MEMO”  button to display the next sets of 
     values. Past readings are displayed from the most 
     recent to the oldest.
4.  This can stored up to 40
 
sets of past readings for each 
     user (4 x 40).
4.2  View Average Readings
1.  Press the “     AVG” button to enter Average function.
2.  If the User Page is not what you want to view, press the 
    
 “PAGE    ” button to select.
3.  Press the “     AVG” button to view  the Average Value of 
     the last 3 stored measuremenets.
To delete an individual measurement from the 
memory,
1.  first press the “PAGE     ” button to select the user.
2.   Press the “MEMO    ” button to select the stored 
      measurement to be deleted from the memory.  
      Press and hold the “     AVG” button for 5 seconds 
      or until “dEL” appears in the display.  Press the 
      “MEMO    ” button to delete.
QUICK  START  GUIDE
 -  Do not use old and new batteries together. 
 -  Align the polarities of batteries correctly.
 -  When the unit will not be used for more than 3 months, 
      remove the batteries. Otherwise, batteries may leak and 
      cause damage to the unit.
Truly Automatic Wrist Bluetooth Blood Pressure Monitor 
DW702 intended to measure the systolic and diastolic 
blood pressure and pulse rate of an adult individual by 
using a non-invasive technique in which an inflatable cuff is 
wrapped around the Wrist.
The devices’ features include irregular pulse rhythm 
detection during measurement, and will display a alert 
signal with the reading when irregular heartbeat is detected.
The devices’ feature include Bluetooth function to transmit 
data to an external Bluetooth device with wireless 
communication .. 
Over-The-Counter Use
3
SAFETY  PRECAUTIONS
2
4
5
6
7
Indications for use
Rev.01
2017/01/06
Manufacturer:    Truly Instrument Ltd.
2/F, Chung Shun Knitting Centre,
                                         1-3 Wing Yip St., Kwai Chung, N.T.,
                                         Hong Kong
TEL:           (852) 24879803
FAX:           (852) 21465756
EC-Representative:Wellkang Ltd t/a Wellkang Tech Consulting
                                         Suite B, 29 Harley Street, LONDON W1G 9QR, 
                                         United Kingdom
22
23
24
25
STORAGE  &  MAINTENANCE
Use a dry, soft cloth to clean the unit, or if desired, use a cloth 
lightly dampened with water.
Keep away the unit from direct sunlight, extreme 
temperatures, humidity or moisture.
Do not use alcohol, benzene, thinner or other volatile liquids 
to clean the unit.
Do not wash or expose the wrist cuff to liquid.
The user must check that the equipment functions safely 
and see that it is in proper working condition before being 
used.
Note: Subject to modification without prior notice.
Model No.
DW702
Measurement Method
Oscillometric
SPECIFICATIONS
EMC Declaration
Guidance and manufacturer’s declaration – electromagnetic immunity
The “DW702” is intended for use in the electromagnetic environment specified below. The customer or the user of the 
“DW702” should ensure that it is used in such an environment. 
Immunity test 
IEC 60601 
test level 
Compliance 
level 
Electromagnetic environment 
guidance 
Electrostatic
discharge (ESD) 
IEC 61000-4-2
±
6 kV contact 
±
8 kV air
±
6 kV contact 
±
8 kV air
Floors should be wood, concrete or 
ceramic tile. If floors are covered with 
synthetic material, the relative humidity 
should be at least 30 %.
Electrical fast 
transient/burst
IEC 61000-4-4 
±
2 kV for power 
supply lines 
±
1 kV for input/output 
lines
±
2 kV for power 
supply lines 
±
1 kV for input/output 
lines
Mains power quality should be that of a 
typical commercial or hospital 
environment.
Surge
IEC 61000-4-5 
±
1 kV differential 
mode
±
2 kV common mode
±
1 kV differential 
mode
±
2 kV common mode
Mains power quality should be that of a 
typical commercial or hospital 
environment.
Voltage dips, short 
interruptions and 
voltage variations 
on power supply 
input lines 
IEC 61000-4-11 
<5 % UT
(>95 % dip in UT)
for 0,5 cycle 
40 % UT
(60 % dip in UT)
for 5 cycles 
70 % UT
(30 % dip in UT)
for 25 cycles 
<5 % UT
(>95 % dip in UT)
for 5 sec
<5 % UT
(>95 % dip in UT)
for 0,5 cycle 
40 % UT
(60 % dip in UT)
for 5 cycles 
70 % UT
(30 % dip in UT)
for 25 cycles 
<5 % UT
(>95 % dip in UT)
for 5 sec
Mains power quality should be that of a 
typical commercial or hospital 
environment. If the user of the “DW702” 
requires continued operation during power
mains interruptions, it is recommended 
that the “DW702” be powered from an 
uninterruptible power supply or a battery. 
Power frequency 
(50/60 Hz) 
magnetic field 
IEC 61000-4-8 
3 A/m
3 A/m
Power frequency magnetic fields 
should be at levels characteristic of a 
typical location in a typical commercial 
or hospital environment. 
NOTE UT is the a.c. mains voltage prior to application of the test level. 
Guidance and manufacturer’s declaration – electromagnetic immunity
The “DW702” is intended for use in the electromagnetic environment specified below. The customer or the user of the 
“DW702” should ensure that it is used in such an environment. 
Immunity test 
IEC 60601 test level 
Compliance
level 
Electromagnetic environment 
– guidance 
Portable and mobile RF communications equipment 
should be used no closer to any part of the “DW702”, 
including cables, than the recommended separation 
distance calculated from the equation applicable to the 
frequency of the transmitter. 
Recommended separation distance
Conducted RF 
IEC 61000-4-6
3 Vrms 
150 kHz to 80 MHz 
3 V
d=1.2P
d=1.2√
P   80MHz to 800MHz 
d=2.3 P  800MHz to 2.5 GHz
Radiated RF 
IEC 61000-4-3
3 V/m 
80 MHz to 2,5 GHz
3 V/m
where is the maximum output power rating of the 
transmitter in watts (W) according to the transmitter 
manufacturer and is the recommended separation 
distance in metres (m). 
Field strengths from fixed RF transmitters, as 
determined
by an electromagnetic site survey,
a
 should be 
less than the compliance level in each frequency 
range.
b
Interference may occur in the vicinity of equipment 
marked with the following symbol: 
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. 
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption 
and reflection from structures, objects and people.
a
b
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land 
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically 
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site 
survey should be considered. If the measured field strength in the location in which the “DW702” is 
used exceeds the applicable RF compliance level above, the Medical DW702 should be observed to verify 
normal  operation.  If  abnormal  performance  is  observed,  additional  measures  may  be  necessary,  such  as 
reorienting or relocating the “DW702”. 
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V
1
] V/m. 
26
27
28
At the end of its use don’t dispose the appliance, including 
removable parts and Accessories, together with the other 
urban waste, but conformably to 2002/96/EC。Since it has 
to be treated apart from home waste,you must take it to
 a differential collection centre specific for electric and 
electronic equipment。Otherwise you can give it back to 
the retailer as you buy an equivalent apparatus。There will 
be Server sanctions in case of transgression。
The batteries used in this device must be disposed of in 
the special bins at the end of their life。
Manufactured in accordance with International Standards
Refer to instruction manual
Keep dry
Type BF applied part
DISPOSAL: Do not dispose this product as unsorted municipal waste. 
                      Collection of such waste separately for special treatment is 
                      necessary.
The european Union flag    
FCC ID:2ABG7-DW702
General warning (not protection against the ingress of water)
The represent of European Union
Disposal
Guidance and manufacturer’s declaration – electromagnetic emissions
The “DW702” is intended for use in the electromagnetic environment specified below. The customer or the 
user of the “DW702” should ensure that it is used in such an environment. 
Emissions test 
Compliance 
Electromagnetic environment – guidance 
RF emissions 
CISPR 11
Group 1
The “DW702” uses RF energy only for its internal 
function. Therefore, its RF emissions are very low and are 
not likely to cause any interference in nearby electronic 
equipment.
RF emissions 
CISPR 11
Class B
Harmonic emissions 
IEC 61000-3-2
Class A
Voltage fluctuations/ 
flicker emissions 
IEC 61000-3-3
Complies
The “DW702” is suitable for use in all establishments, 
including domestic establishments and those directly 
connected to the public low-voltage power supply network 
that supplies buildings used for domestic purposes. 
Recommended separation distances between 
portable and mobile RF communications equipment and the Medical DW702
The “DW702” is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. 
The customer or the user of the Medical DW702 can help prevent electromagnetic interference by maintaining a 
minimum distance between portable and mobile RF communications equipment (transmitters) and the “DW702” as 
recommended below, according to the maximum output power of the communications equipment. 
Separation distance according to frequency of transmitter 
m
Rated maximum output 
power of transmitter 
W
150 kHz to 80 MHz 
80 MHz to 800 MHz 
800 MHz to 2,5 GHz 
0,01
0.12
0.12
0.23
0,1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance in
meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where is the 
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. 
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. 
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and 
reflection from structures, objects and people.
Automatic Wrist Bluetooth Blood 
Pressure Monitor
DW702
Instruction Manual
FCC Statement 
This equipment has been tested and found to comply with 
the limits for a Class B digital device, pursuant to part 15 of 
FCC Rules. These limits are designed to provide reasonable 
protection against harmful interference in a residential 
installation. This equipment generates and can radiate 
radio frequency energy and, if not installed and used in 
accordance with the instructions, may cause harmful 
interference to radio communications. However,there is 
no guarantee that interference will not occur in a particular 
installation. If this equipment does cause harmful 
interference to radio or television reception,which can 
be determined by turning the equipment off and on, the 
user is encouraged to try to correct the interference by one 
or more of the following measures: 
Attention Don’t disassemblerepair and renovation 
                             ontology
When the screen display the battery for one degreeit means
that low Voltage and must change the new one to void to
measure error
- The monitor uses the oscillometric method to measure 
  systolic and diastolic blood pressure, as well as heart rate;
- This device is intended for use in measuring blood 
  pressure and pulse rate in the adult population, do not use 
  this device on infants or persons who cannot express their 
  intentions;
  The monitor is not intended to be a diagnostic device, It is 
  a home healthcare product only and it is not intended to 
  serve as a substitute for the advice of a physician or 
  medical professional.
- Don't not use this device for diagnosis or treatment of any 
  health problem or disease, Measurement results are for 
  reference only, consult a healthcare professional for 
  interpretation of pressure measurements, Contact your 
  physician if you have or suspect any medical problem. Do 
  not change your medications without the advice of your 
  physician or healthcare professional;
- Use only Truly authorized parts and accessories. Parts and 
  accessories not approved for use with the device may 
  damage the unit. and get a error of measurement;
- Proper cuff size is critical for accurate measurements, 
  Follow the instructions in this manual and printed on the 
  arm cuff to ensure the appropriate size of cuff is being 
  used; 
- Blood pressure measurement can be affected by the 
  position of the user,his or her physiologic condition and 
  other factors;Position of the user such as bending over the 
  body or cross-legged sitting will produce abdominal 
  pressure or make the position of the arm below the heart, 
  that will increase the blood pressure;
- This product is not suitable for people with arrhythmias or 
  serious arteriosclerosis. It may have difficulty determining 
  the proper blood pressure for pregnant women and for 
  users with irregular heartbeat,diabetes, pre-eclampsia, 
  poor circulation of blood, measurement on the arm on the 
  side of a mastectomy,kidney problems or for users who 
  have suffered from a stroke;
- When measurement,The pressurization of the cuff to arm 
  where intravascular access or therapy, or an arterio-
  venous(A-V) shunt, is present because of temporary 
  interference to blood flow  and could result in injury to the 
  patient;
- When measurement, The pressurization of the wrist can 
  temporarily cause function of simultaneously used 
  monitoring other device on the same limb;
- The pulse display is not suitable for checking the 
  frequency of heart pacemakers;
- Do not use a cellular phone near the device, It may result 
   in an operational failure;
- Do not use the monitor in the driving or flying vehicles;
- Frequenly measuring will make the blood not flow, cause 
  the arm-numbness and abnormal reaction of body;
- When you are using the monitor to measure especially 
  frequently used, the arm is compressed by the cuff, and 
  then the fingertip will be caused congestion. During the 
  congestion, please loosen the cuff and lift the hands over 
  your head, and make your right and left hand squeese and 
  stretch 15 times, them will unclog the congestion;
- Do not applying the CUFF over a wound, as this can cause 
   further injury;
- When measurement, please check that operation of the 
  device does not result in prolonged impairment of the 
  circulation of the blood of the patient;
- Do not compress the cuff tube during the measurement, 
  or it will cause fail inflation or affect the result of 
  measurement;
- Please use the product in the approved operate 
  environment, or will cause the inaccurate result of 
  Handling batteries properly: measurement;
- Dispose of the device, components and optional 
  accessories according to applicable local regulations, 
  Unlawful disposal may cause environmental pollution.
Don’t calling around the Blood pressure monitor when it is 
using 
,Don’t move and avoid electromagnetic interference 
and noise interference when measuring that to avoid to 
measure error
- Babies, young children or individuals who cannot express 
  their consent are not suitable to take blood pressure 
  measurement.
- Blood pressure readings may change in case of 
  pregnancy. Pregnant women can consult their doctor 
  before taking measurement.
- Individuals with serious arteriosclerosis are not suitable 
   to take blood pressure measurement.
- Keep the equipment away from children and pets.
- Do not disassemble, repair and alteration body.
- Self-measurement is not medical treatment. If there are 
   unusual values, please consult your doctor.
- Handling batteries properly:
   -  As soon as old batteries run out, replace with new 
      batteries.
Avoid smoking, eating or exercising for 30 minutes before 
taking measurement
8
1.  Sit upright in a chair with your both 
     feet on the floor.
3.  Pull on the end of the cuff until it wraps 
     securely around your wrist. Place your 
     arm on a table so that the cuff will be 
     at the same level as your heart.
4.  Connect iOS device via Bluetooth (Turn Bluetooth "On" 
     under the "Settings" menu on the iOS device).
5.  Open "TRULY" application, and then press the Blood 
     Pressure icon            on Measure screen.
6.  Press the “     ” button. The cuff will automatically inflate and 
     the measurement will start.
7.  In the same time of inflating, The monitor automatic pair 
    quickly to TRULY application in a few seconds. Once 
    monitor connect to the Application, the color of        on 
    the iOS application screen will be change to green from 
    gray.
2.  Remove tight fitting clothing from you 
     upper arm and thick clothing.
Setup Procedures Installation
Prior to first use, Please download the TrulyHealth application
from Apple store and install it to your iOS device.
(Bluetooth 4.0 capabilities, e.g. iPhone 4S, iPhone 5, NewiPad 
and iPadMini).
Installing the APP
1) Download and install the Truly Healthcare App onto your 
     Bluetooth 4.0 supported smart devices.  The free Truly 
     Healthcare App is available on the App Store for the 
     Apple devices and Google Play for the Android based 
     devices.
2) Open the App on your smart device and follow the on 
     screen instructions to set up the user profile.
Warning:
The App is only intended use for receiving the measurement 
result from the Blood Pressure Monitor. It can't control the 
monitor.
9
11
2.1  Battery Installation
3.  Slide the battery cover to close 
     until the lock securely clicks into 
     place.
PREPARATION
1.  Remove the batteries cover in 
     according with the direction 
     of arrow.
2.  Remove the used batteries and 
     insert 2 AAA new batteries into 
     the battery compartment an 
     shown,taking care that the 
     polarities (+)and (-) match the 
     polarities of the battery 
     compartment as indicated.
12
2.2  Date & Time Setting
Note:
   -    Set the correct date and time before taking a 
         measurement.
   -    Date and time should be reset after replacing new 
         batteries.
13
1.  Press and hold the “     AVG”  button for a few seconds until 
     the Year display flashes.
2.  Press the“   MEMO” or “PAGE    ”  button to select the current 
     year.
3.  Press the“     AVG”   button to confirm Year Setting.
4.  Follow the same steps to set the Month, Date, Hour and 
     Minute.
Noted: If you select Clock Function “ON”, you will continue at 
              the Alarm Setting section. Otherwise, setting will end.
5.  You can choose Clock Function “ON” or “OFF” while the 
     monitor is not in use. Press “   MEMO” or  “PAGE    ”  button to 
     select. Press the  “     AVG” button to confirm.
6.  To disable Alarm, Press the ”    MEMO ” or  “PAGE    ”  button 
     to select “OFF” and press the ”     AVG”  button to confirm.
7.  To set 1st Alarm, Press the”    MEMO”  or ” PAGE    ”  button to 
     select “ON”.  Press the ”     AVG”  button to confirm.
8.  When Hour digit flashes, press the“   MEMO” or  “PAGE    ” 
     button to select Hour.  Press the“     AVG”  button to 
     confirm.
9.  When Minute digit flashes, press the“   MEMO” or  “PAGE    ” 
     button to select Minute.  Press the“     AVG”  button to 
     confirm.
10. Follow the same steps to set the 2nd, 3rd, and 4th Alarm. 
Note:
   -    Press the “     ” button anytime during clock setting/alarm 
         setting to start over if an error occurs. 
   -    Each alarm clock lasts 45 seconds, or you can turn it off 
         by pressing the “     ” button.
2.3  Alarm Setting(This alarm is only tips for patient)
Note:
   -    You will not be able to set alarm clock if Clock Function 
         is set to “OFF”.
   -    You can set a maximum of 4 alarms.
14
2.4  Language Setting
1.  Press and hold the “M” button until the screen displays 
     [LLLx] (x = 0 ~ 3), then release the button. 
2.  Press the “M” button to select language. Number 0 ~ 3 
     represents the following languages:
     0   Mute
     1   English
     2   French
     3   Spanish
3.  Press the “     ” button to confirm the selection and begin 
     measurement.
Note:Replace batteries when the battery indicator “       ” 
              shows low voltage.
To Change the Language Format:
When a measurement is completed, the results will read 
aloud; the systolic value, diastolic value and pulse.The 
default language is English.
10
Monitor
Features
OVERVIEW
Memory Recall
          (Up)
Average Values
          (Set)
Page Select
        (Down)
Power
(Start)
WHO Light
Indictor
Date & Time
User 1~ 4
Memory Average
Systolic
Diastolic
Low Battery
Indicator
Pulse Symbol
Memory No.
(1-40 readings)
•  Operated by 2 AAA batteries
•  Auto-off Power Saving Feature
•  Averaging Mode -The average value of the last 3 
   measurements stored in the memory will display
•  4 Alarm Clock Settings
•  4 User / 160 Total Memory
•  Blood Pressure Mode and Clock Mode   
•  Irregular Heartbeat Detector   
•  One Touch Operation
Memory
When using more than one monitor with the Application, 
make sure that the WHO Light(      ) in the subject monitor will 
turn on and in Green during inflation.  
This indicates that the monitor is successfully connected 
to the Application.  If the WHO Light does not light up in 
Green on the subject monitor,  your monitor is not connected 
to the Application. In this case, turn off the monitor with the 
Green WHO Light and go back to Step 6 again.
8.  When the measurement is complete, 
     the cuff will automatically deflate 
     and your systolic and diastolic 
     pressure values and pulse rate 
     will display.
9.  After that check the measurement 
     history on History screen or chart 
     screen by iOS application.
Warning:
1. After press the "     " button, the color of  "        " on the iOS 
    application screen is always gray . It means Monitor can 
    connect to the application, please repeat  the operation 
    step 4~6.
2. In the same time, one monitor only pair and connect to 
    one application via Bluetooth.When there are more than 
    one compatible device in the vicinity of the subject 
    device.  
The device only connect to the subject device of first 
turning on . if you want connecting other subject device, 
you should first turn off the current connecting device.
Measurement Range
Pressure: 20 – 280 mmHg
Pulse Rate: 40 – 195 beats/min
Accuracy
Pressure: +/- 3 mmHg
Pulse Rate: +/- 5%
Inflation
Pump driven
Pressure Detection
Semi-conductor
Power Supply
2x AAA batteries 
Auto Power Off
3 minutes after last operation
Storage Environment
-20°C to 60°C, 10% to 95% RH
Operation Environment 10°C to 40°C, 30% to 85% RH
 70kPa to 106kPa
External Dimensions
80 x 85 x 60 mm
Net Weight
111g (batteries excluded)
Cuff Dimensions
Wrist circumference of 135~220mm
Package Content
Instruction Manual, Battery Set, 
Warranty Card
3.15” (L) x 3.35”(W) x 2.36” (H) or 
used life span of battery
The pressure of the 
running
70KPa – 106KPa
1000 times
V055
Software version
Application part
(out of contact the body)
    connect tub、shell
(contact the body) cuff
Reorient or relocate the receiving antenna。 
Increase the separation between the equipment and 
receiver。
Connect the equipment into an outlet on a circuit different 
from that to which the receiver is connected。
Consult the dealer or an experienced radio/TV technician 
for help。
This device complies with Part 15 of FCC Rules。Operation 
is subject to the following two conditions:(1)This device 
may not cause harmful interference, and  (2)This device 
must accept any interference received, including 
interference that may cause undesired operation。
Note:The manufacturer is not responsible for any radio 
              or TV interference caused by unauthorized 
              modifications to this equipment. such modifications 
              could void the user’s authority to operate this 
              equipment。
Warning: Long power cable and air hose. To avoid 
                    strangulation and entanglement, keep cable and 
                    hoses out of reach of young children.