Innokas Yhtyma Oy VC150 Manuale Utente
5-2
VC150 Vital Signs Monitor
KO00065K
Patient and caregiver data: Description
Description
The Patient screen allows you to access stored patient data. Retrieving
snapshot points is especially useful when doing hospital rounds: if the patient’s
temperature and SpO
snapshot points is especially useful when doing hospital rounds: if the patient’s
temperature and SpO
2
measurements are taken while an NIBP determination is
in progress, then upon completion of the determination, the vital signs for a
patient can be analyzed on screen or printed out.
patient can be analyzed on screen or printed out.
The following information refers to operation in clinical mode. The monitor can
hold 3000 stored entries in history. It displays the most recent entries first.
Entries are automatically removed when they become older than 24 hours if
these are not printed to a PDF or queued up for sending to the EMR.
hold 3000 stored entries in history. It displays the most recent entries first.
Entries are automatically removed when they become older than 24 hours if
these are not printed to a PDF or queued up for sending to the EMR.
The patient data is maintained and displayed in the Patient screen with a time
stamp at the top of the snapshot. Missing parameter data leaves an empty
column in the table. Admitting a new patient also creates an empty column. A
new snapshot is added to the left side of the table.
stamp at the top of the snapshot. Missing parameter data leaves an empty
column in the table. Admitting a new patient also creates an empty column. A
new snapshot is added to the left side of the table.
An entry is stored in local clinical history at the completion of an NIBP
determination and at the completion of an Exergen temperature measurement
or a successful predictive Welch Allyn temperature measurement. At the end of
an NIBP determination, systolic, diastolic, MAP and PR values are stored.
determination and at the completion of an Exergen temperature measurement
or a successful predictive Welch Allyn temperature measurement. At the end of
an NIBP determination, systolic, diastolic, MAP and PR values are stored.
NOTES
When SpO
2
is being monitored continuously, values are only stored when
the snapshot icon is selected. Predictive temperature values are stored
separately after each determination. If Snapshot upon NIBP completion
has been selected in configuration mode, then current SpO
separately after each determination. If Snapshot upon NIBP completion
has been selected in configuration mode, then current SpO
2
value is stored
when NIBP measurement is complete.
Only those items that are selected to display on the screen are stored in the
snapshot. Visible screen items can be adjusted in the related parameter tab
in Monitor Setup.
snapshot. Visible screen items can be adjusted in the related parameter tab
in Monitor Setup.
CAUTION
If you need to use another unit of measurement for NIBP, first send all
important patient data to the hospital EMR and then ask service to
change the unit of measurement.
important patient data to the hospital EMR and then ask service to
change the unit of measurement.
Stored entry
field for a
patient
field for a
patient
Patient list