Chapter 4 - Operation
32
Standard Operation
This section describes the standard recommended procedure to follow when preparing
to detect drives and mount drive’s volume in a write-protected environment using Super
DriveLock unit.
1. Verify Installation
to detect drives and mount drive’s volume in a write-protected environment using Super
DriveLock unit.
1. Verify Installation
Verify that the Super DriveLock and PC/Notebook are properly configured using
the Installation instructions listed in
2. Prepare the Drive for Detection and Mounting
Verify that the Suspect’s drive jumper block is properly configured.
For P-ATA drives the jumper block should be set for “Single/Master” operation.
For SCSI drives the SCSI Address jumper block should be set to 0.
For SCSI drives the SCSI Address jumper block should be set to 0.
Using the supplied data and power cable(s), connect the drive to the SDL’s
appropriately labeled interface connector. Connectors are available for P-ATA,
S-ATA, SCSI and USB devices. Refer to
S-ATA, SCSI and USB devices. Refer to
display of drive positions.
For P-ATA data cables, connect the cable’s “HDD Side” labeled connector to the
drive’s data connector. Verify that the cable’s Pin-1 is properly oriented. The
cable’s Pin-1 should be aligned adjacent to the drive’s power connector
(See Fig. 23).
cable’s Pin-1 should be aligned adjacent to the drive’s power connector
(See Fig. 23).
For SCSI data cables, connect the ICS supplied single channel SCSI cable to the
unit’s SCSI data connector. Verify that the SCSI Cable Terminator is connected
to the cable’s end connector. The terminated connector is located on the drive
side (See Fig. 24).
to the cable’s end connector. The terminated connector is located on the drive
side (See Fig. 24).
NOTE: A short and long S-ATA and P-ATA data and power cable set is provided
for ease of use.