Nonin Medical Onyx II 9560 6343-002 Manuale Utente
Codici prodotto
6343-002
Your Partner of Choice.
With more than 20 years of experience and dedication to the design and support of
noninvasive monitoring devices, NONIN has helped many medical professionals meet clinical and economic objectives.
Trusted by clinicians worldwide, countless tests are performed every day using NONIN pulse oximeters in more than
125 countries. More than 90 OEM partners around the world also rely on NONIN’s digital technology to successfully
integrate superior pulse oximetry and capnography into their products. NONIN’s dedication to technological leadership,
precision manufacturing and uncompromised customer support ensures quality products and service you expect.
Trusted by clinicians worldwide, countless tests are performed every day using NONIN pulse oximeters in more than
125 countries. More than 90 OEM partners around the world also rely on NONIN’s digital technology to successfully
integrate superior pulse oximetry and capnography into their products. NONIN’s dedication to technological leadership,
precision manufacturing and uncompromised customer support ensures quality products and service you expect.
®
www.nonin.com
Nonin Medical, Inc.
13700 1st Avenue North
Phone: +1.763.553.9968
Plymouth MN 55441-5443, USA
Fax:
+1.763.553.7807
Email: info@nonin.com
Toll Free: 800.356.8874
Specifications are subject to change without notice.
©2008 Nonin Medical, Inc.
All trademarks are the property of Nonin Medical, Inc unless otherwise noted.
Bluetooth
All trademarks are the property of Nonin Medical, Inc unless otherwise noted.
Bluetooth
®
and the Bluetooth Logo
®
are registered trademarks of Bluetooth SIG, Inc.
6459-001-01
Oxygen Saturation Display Range
0 – 100% SpO
2
Pulse Rate Display Range
18 to 321 beats per minute (BPM)
Oxygen Saturation Declared Accuracy Range (A
rms
*) 70 – 100% SpO
2
±2 digits
Low Perfusion Oxygen Saturation Declared
Accuracy Range (A
Accuracy Range (A
rms
*)
70 – 100% SpO
2
±2 digits
Pulse Rate Declared Accuracy Range (A
rms
*)
20 – 250 BPM ±3 digits
Low Perfusion Pulse Rate Declared Accuracy
Range (A
Range (A
rms
*)
40 – 240 ±3 digits
Measurement Wavelengths and Output Power**
Red:
660 nanometers @ 0.8 mw maximum average
Infrared:
910 nanometers @ 1.2 mw maximum average
Temperature
Operating:
+41° to +104°F / 5° to +40°C
Storage/Transportation:
-40° to +158°F / -40° to +70°C
Humidity
Operating:
10 – 95% noncondensing
Storage/Transportation:
10 – 95% noncondensing
Altitude
Operating:
Up to 40,000 feet / 12,192 meters
Hyperbaric Pressure:
Up to 4 atmospheres
Battery Life
Operating:
Approximately 600 spot checks
Storage:
1 year, minimum
Antenna
Type:
L-shaped PWB whip-type antenna
Transmitter
Bluetooth Compliance:
Version 2.0
Operating Frequency:
2.4 – 2.4835 GHz
Output Power:
<20 dBm
Operating Range:
100-meter radius indoors***
Network Topology:
Star
Operation:
Bluetooth Slave
Antenna Type:
Internal
Modulation Type:
Gaussian Frequency Shift Keying (GFSK)
Frequency Hopping Spread Spectrum
Band Width:
1 MHz
Classifications per IEC 60601-1 / CSA 601.1 / UL 60601-1
Degree of Protection:
Type BF-Applied Part
Enclosure Degree of Ingress Protection:
IP32
Mode of Operation:
Continuous
This product complies with ISO 10993-1.
*± 1 Arms represents approximately 68% of measurements.
**This information is especially useful for clinicians performing photodynamic therapy.
***Line of sight when connected to a Class I device.
**This information is especially useful for clinicians performing photodynamic therapy.
***Line of sight when connected to a Class I device.
Specifications