:TRIGger:SLOPe:WHEN
Syntax
:TRIGger:SLOPe:WHEN <when>
:TRIGger:SLOPe:WHEN?
:TRIGger:SLOPe:WHEN?
Description
Set or query the trigger condition in slope trigger.
Parameter
Name
Type
Range
Default
<when> Discrete {PGReater|PLESs|NGReater|NLESs|PGLess|NGLess} PGReater
Explanation
).
The oscilloscope triggers when the positive slope time of the input signal is
greater/lower than the specified time.
).
The oscilloscope triggers when the negative slope time of the input signal is
greater/lower than the specified time.
) of time. The oscilloscope
triggers when the positive/negative slope time of the input signal is greater than the
specified lower limit and lower than the specified upper limit.
Return
Format
The query returns PGR, PLES, NGR, NLES, PGL or NGL.
Example
:TRIGger:SLOPe:WHEN NLESs /*Set the trigger condition to NLESs*/
Related
Command
:TRIGger:SLOPe:TIME
Syntax
:TRIGger:SLOPe:TIME <time>
:TRIGger:SLOPe:TIME?
:TRIGger:SLOPe:TIME?
Description
Set or query the time value in slope trigger. The default unit is s.
Parameter
Name
Type
Range
Default
<time>
Real
8ns to 10s
PGReater, NGReater: 1μs
PLESs, NLESs: 2μs
Explanation
) is
PGReater, PLESs, NGReater and NLESs.
Return
Format
The query returns the time value in scientific notation.
Example
:TRIGger:SLOPe:TIME 0.000003 /*Set the time value to 3μs*/
:TRIGger:SLOPe:TIME? /*The query returns 3.000000e-06*/
Related