Lecroy WAVEACE 2004 4-channel oscilloscope, Digital Storage oscilloscope, WaveAce 2004 Manuale Utente

Codici prodotto
WaveAce 2004
Pagina di 89
Operator's Manual 
920836 Rev A  
47 
MISCELLANEOUS – AUTO-CALIBRATE, ACAL 
Description 
Available only on WaveAce 2000 models, AUTO_CALIBRATE enables or disables automatic 
calibration of the oscilloscope. If ACAL is ON at power-up, all input channels are periodically 
calibrated for the current input amplifier and timebase settings, whether the instrument has 
been adjusted or not.  This is separate from the calibration the instrument performs whenever 
you adjust a gain or offset. 
ACAL OFF disables Auto-calibration. You can issue a *CAL? query at any time to fully calibrate 
the oscilloscope, but periodic calibrations stop if ACAL is OFF. 
The response to the AUTO_CALIBRATE? query indicates whether auto-calibration is enabled or 
disabled. 
Command Syntax 
Auto_CALibrate <state> 
<state>:= {ON, OFF} 
Query Syntax 
Auto_CALibrate? 
Response Format 
Auto_CALibrate <state> 
Related Commands 
*CAL? 
MISCELLANEOUS - BUZZER, BUZZ 
Description 
The buzzer command controls the built-in buzzer. By means of the BEEP argument, the buzzer 
can be activated to sound short beeps at key points in the control program.   
Command Syntax 
BUZZer <state> 
<state>:=  {ON, OFF} 
Example 
The following instruction turns off the buzzer: 
BUZZ OFF