Lecroy WAVEACE 2004 4-channel oscilloscope, Digital Storage oscilloscope, WaveAce 2004 Manuale Utente

Codici prodotto
WaveAce 2004
Pagina di 89
Operator's Manual 
920836 Rev A  
Remote Control Through LAN   
WaveAce 2000 series oscilloscopes have a standard LAN connection port for remote control 
that utilizes Teledyne LeCroy's VICP protocol for transmitting messages. This protocol emulates 
IEEE 488.2 (GPIB) and includes operation bits corresponding to SRQ, EOI, Clear, and others in a 
header that is defined by the VICP protocol.  
The WaveAce must be assigned a static IPv4 address for remote control; do not use an address 
from the DHCP pool. 
You can send remote commands to the oscilloscope directly from the WaveStudio® terminal.  
WaveStudio can be downloaded free at www.teledynelecroy.com/support/softwaredownload, 
under Oscilloscope Software Utilities. 
On the Controller 
For communication using the VICP (LAN) interface, no additional driver software is required 
when using WaveStudio. For VICP communication with other applications that use the VISA 
standard, NI-VISA must be installed, along with Teledyne LeCroy's 'VICP Passport'. The Passport 
extends VISA to support Teledyne LeCroy's VICP protocol. The VICP passport may be 
downloaded from www.teledynelecroy.com. Also see the application brief LAB_WM827 
Understanding VICP and the VICP Passport on Teledyne LeCroy's website for more information. 
On the Instrument 
1.  Attach an Ethernet cable from the LAN port on the rear of the scope to an Ethernet port on 
the controller machine or a connection device on the same network as the controller. 
 
2.  From the WaveAce menu, choose Utility > IP Settings (on the third page of settings).