Lecroy WAVEACE 2002 2-channel oscilloscope, Digital Storage oscilloscope, WaveAce 2002 Manuale Utente
Codici prodotto
WaveAce 2002
Operator's Manual
106
WA1K2K-OM-E RevB
(3 V/m, 80-1000 MHz; 3 V/m, 1400 MHz - 2 GHz; 1 V/m, 2 GHz -
2.7 GHz)
2.7 GHz)
EN 61000-4-4:2004** Electrical Fast Transient/Burst.
(1 kV on power supply lines, 0.5 kV on I/O signal data and control
lines)
lines)
EN 61000-4-5:2006** Surges.
(1 kV AC Mains, L-N, L-PE, N-PE)
EN 61000-4-6:2007** RF Conducted Electromagnetic Field.
(3 Vrms, 0.15 MHz - 80 MHz)
EN 61000-4-11:2004† Mains Dips and Interruptions.
(1 cycle voltage dip, 100% short interruption)
* In order to conform to Radiated Emissions standards, use properly
shielded cables on all I/O terminals.
shielded cables on all I/O terminals.
** Meets Performance Criteria "B" limits during the disturbance; product
undergoes a temporary degradation or loss of function of performance
which is self recoverable.
undergoes a temporary degradation or loss of function of performance
which is self recoverable.
† Meets Performance Criteria "C" limits during the disturbance; product
undergoes a temporary degradation or loss of function of performance
which requires operator intervention or system reset.
undergoes a temporary degradation or loss of function of performance
which requires operator intervention or system reset.
WARNING
This is a Class A product. In a domestic environment this product may
cause radio interference, in which case the user may be required to take
appropriate measures.
cause radio interference, in which case the user may be required to take
appropriate measures.
Low-Voltage Directive
EN 61010-1:2001
Safety requirements for electrical equipment for measurement, control,
and laboratory use.
and laboratory use.
The oscilloscope has been qualified to the following EN 61010-1 limits: