AC Power Source 
100 to 240 VAC (±10%) at 50/60/400 Hz (+10%); Installation Category: 
300V CAT II   
No manual voltage selection is required because the instrument 
automatically adapts to line voltage. 
Power Consumption 
< 50 Watts for all models (depending on accessories installed - probes, 
USB devices, etc.). 
Power consumption in Standby Mode: < 5 Watts. 
Power and Ground Connections 
The instrument is provided with a 10A/250V 18AWG rated grounded 
cord set containing a molded three-terminal polarized plug and a 
standard IEC-60320 (Type C13) connector for making line voltage and 
safety ground connections. The AC inlet ground terminal is connected 
directly to the frame of the instrument. For adequate protection against 
electrical shock hazard, the power cord plug must be inserted into a 
mating AC outlet containing a safety ground contact. 
WARNING - Electrical Shock Hazard 
Only use the power cord provided with your instrument. 
Any interruption of the protective conductor inside or outside of the 
oscilloscope, or disconnection of the safety ground terminal creates a 
hazardous situation. Intentional interruption is prohibited. 
In Standby mode, the oscilloscope is still connected to the AC supply. The 
instrument can only be placed in a complete Power Off state by flipping 
the main power switch on the rear of the instrument (if provided) to the 
off (Zero) position or by removing the power cord. In cases where no 
main power switch is provided, removing the main power cord is the 
disconnect device.