Lecroy WAVEACE 2002 2-channel oscilloscope, Digital Storage oscilloscope, WaveAce 2002 Manuale Utente

Codici prodotto
WaveAce 2002
Pagina di 89
WaveAce Remote Control 
 
920836 Rev A  
Character Data 
These are simple alphanumeric words or abbreviations indicating a specific action. 
In commands where you can specify many parameters, or where not all parameters are 
applicable at the same time, the format requires pairs of character data values. The first value 
names the parameter to be modified, while the second gives its value.  
HARDCOPY_SETUP DEST,PRINTER,PRINTER,EPSON 
Here, two pairs are shown: the first specifies the DESTination is a PRINTER, while the second 
specifies the PRINTER is EPSON. Any HARDCOPY_SETUP parameters that are not relevant for 
printers, or are left unchanged, are omitted from the command. 
Numeric Data 
The numeric data type is used to enter quantitative information. Numbers can be entered as 
integers, fractions, or exponents: 
C1:VPOS -5
 
C2:OFST 3.56
  
TDIV 5.0E-6
 
Numeric values can be followed with multipliers and units modifying the value of the numeric 
expression. The following table of mnemonics is recognized: 
Multiplier 
Exponential Notation 
Suffix 
EX 
1E18 
Exa- 
1E12 
Tera- 
MA 
1E6 
Mega- 
1E-3 
milli- 
1E-9 
nano- 
1E-15 
femto- 
PE 
1E15 
Peta- 
1E9 
Giga- 
1E3 
kilo- 
1E-6 
micro- 
1E-12 
pico- 
1E-18 
atto- 
String Data 
This is used to send multiple characters as a single parameter. Enclose any sequence of ASCII 
characters between single or double quotation marks: 
DIRectory DISK,UDSK,ACTION,CREATE,'20120801Results'