Lecroy WAVEACE 2002 2-channel oscilloscope, Digital Storage oscilloscope, WaveAce 2002 Manuale Utente

Codici prodotto
WaveAce 2002
Pagina di 89
WaveAce Remote Control 
 
46 
920836 Rev A  
MISCELLANEOUS - *CAL? 
Description 
The *CAL? query causes the oscilloscope to perform an internal self-calibration and generates a 
response that indicates whether or not your oscilloscope completed the calibration without 
error. This internal calibration sequence is the same as that which occurs at power-up. At the 
end of the calibration, after the response has indicated how the calibration terminated, the 
oscilloscope returns to the state it was in just prior to the calibration cycle. 
Query Syntax 
*CAL? 
Response Format 
*CAL? <diagnostics> 
<diagnostics>:=  0 (calibration successful)  
Response message (if no failure): 
*CAL 0 
MISCELLANEOUS - *IDN? 
Description 
The *IDN? query causes the instrument to identify itself. The response comprises 
manufacturer, oscilloscope model, serial number, and firmware revision level. 
Query Syntax 
*IDN? 
Response Format 
*IDN LECROY,<model>,<serial_number>,<firmware_level> 
<model>:=
 A six- or seven-character model identifier 
<serial_number>:=
 A nine- or 10-digit decimal code 
<firmware_level>:= major release, minor release, and update levels formatted 
xx.y.z 
Example response message: 
*IDN LECROY,WaveAce1012,LCRY2150C12345,5.01.02.09