WAVEFORM TRANSFER - WAVEFORM, WF 
Description 
The WAVEFORM command transfers a waveform from the controller to the oscilloscope.  The  
WAVEFORM? query reads a waveform from the oscilloscope to the controller. 
WAVEFORM stores an external waveform back into the oscilloscope's internal memory. A 
waveform consists of several distinct entities: 
 
DESC  - The descriptor 
 
TEXT  - The user text  
 
TIME  - The time descriptor 
 
DAT1 - The data block 
and, optionally... 
 
DAT2 - A second block of data 
NOTE: Only complete waveforms queried with WAVEFORM? ALL can be restored to the 
oscilloscope. 
The WAVEFORM? query instructs the oscilloscope to transmit a waveform to the controller. The 
entities can be queried independently. When the ALL parameter is specified, all four or five 
entities are transmitted in one block in the order listed above. 
NOTE: The format of the waveform data depends on the current settings specified by 
the last WAVEFORM_SETUP command. 
Command Syntax 
<memory>:WaveForm ALL,<data_block> 
<memory>:= {M1,M2,M3,M4,M5,M6,M7,M8,M9,M10} 
<data_block>:= waveform data block. 
Query Syntax 
<trace>:WaveForm? <entity> 
<trace>:= {TA,TB,TC,TD,C1,C2,C3,C4,M1,M2,M3,M4,M5,M6,M7,M8,M9,M10}  
<entity>:= {DESC, TEXT, TIME, DAT1, DAT2, ALL} 
If you do not provide a parameter, 
ALL
 is assumed.