DISPLAY - HOR_POSITION, HPOS 
Description 
The HOR_POSITION command repositions the geometric center of a zoom to display different 
portions of the trace. Allowed positions range from division 0 through 10. If the source trace 
was acquired in sequence mode, the shift will only apply to a single segment at a time. 
The difference between the specified and the current horizontal position is applied to all 
zooms. However, if the difference would cause any zoom trace to go outside the left or right 
screen boundaries, the difference is adapted before being applied to the traces. 
The HOR_POSITION? query returns the position of the geometric center of the zoom. 
NOTE: Segment number 0 has the special meaning Show All Segments Unexpanded
Command Syntax 
Hor_POSition <hor_position>,<segment> 
<hor_position>:= 0 to 10 DIV 
<segment>:= 0 to max segments 
NOTE: The segment number is only relevant for waveforms acquired in sequence mode; 
it is ignored in single waveform acquisitions. When the segment number is set to 0, all 
segments are shown. The unit DIV is optional. 
Query Syntax 
Hor_POSition? 
Response Format 
Hor_POSition <hor_position>[,<segment>] 
Example 
The following instruction positions the center of the zoom at division 3: 
HPOS 3