Vertical Settings and Channel Controls 
Vertical Settings and Channel Controls 
When you turn a channel trace ON, the Channel menu opens. The 
Channel menu page shown (1/3, 2/3, or 3/3) is always based on the most 
recent trace activated. 
 
Choosing Coupling 
You can choose one of these input coupling modes: 
 
DC - Passes both AC and DC components of the input signal. 
 
AC - Blocks the DC component of the input signal and attenuates 
signals below 10 Hz. 
 
GND - Disconnects the input signal. Use GND coupling to display a 
zero-volt waveform. Internally, the channel input is connected to a 
zero-volt reference level. 
PLEASE NOTE THE FOLLOWING:  
 
 If the channel is set to DC coupling, you can quickly measure the 
DC component of the signal by simply noting its distance from the 
ground symbol. 
 
If the channel is set to AC coupling, the DC component of the 
signal is blocked allowing you to use greater sensitivity to display 
the AC component of the symbol.