Rigol DS1104Z 4-channel oscilloscope, Digital Storage oscilloscope, DS1104Z Manuale Utente
![Rigol](https://files.manualsbrain.com/attachments/4ef153c5b3a9e56ac40ea738ae281ea9dc808cdc/common/fit/150/50/a84f25c70c016e6b8d8e8d0fbe042c3bbe7b8a8feeee0ecbbc25ec79cb57/brand_logo.gif)
Codici prodotto
DS1104Z
RIGOL
DS1000Z User’s Guide
14-9
Self-calibration
The self-calibration program can quickly make the oscilloscope reach the best
working state to get the most precise measurement values. You can perform
self-calibration at any time especially when the change of the environment
temperature is up to or more than 5 ℃. Make sure that the oscilloscope has been
warmed up or operated for more than 30 minutes before the self-calibration.
Disconnect all the input channels and then press Utility Self-Cal and the
self-calibration interface as shown in the figure below is displayed.
Press Start and the oscilloscope will start to execute the self-calibration program.
Press Exit to give up the self-calibration operation at any time and return to the
previous menu.
Note: most of the keys are disabled during the self-calibration.