Rigol DS1104Z 4-channel oscilloscope, Digital Storage oscilloscope, DS1104Z Manuale Utente
Codici prodotto
DS1104Z
RIGOL
DS1000Z User’s Guide
17-1
Chapter 17 Specifications
All the specifications are guaranteed except parameters marked with “Typical” and
the oscilloscope needs to operate for more than 30 minutes under the specified
operation temperature.
Sample
Sample Mode
Real-time sample
Real-time Sample
Rate
Rate
1 GSa/s (single-channel)
500 MSa/s (dual-channel)
250 MSa/s (four-channel)
500 MSa/s (dual-channel)
250 MSa/s (four-channel)
Averaging
After both the channels finish N samples at the same time, N
can be 2, 4, 8, 16, 32, 64, 128, 256, 512, 1024.
can be 2, 4, 8, 16, 32, 64, 128, 256, 512, 1024.
High Resolution
The highest resolution is 12 bit
Peak Detection
4 ns
Memory Depth
Single-channel: Auto, 12k pts, 120k pts, 1.2M pts, 12M pts
and 24M pts (option) are available
Dual-channel: Auto, 6k pts, 60k pts, 600k pts, 6M pts and
12M pts (option) are available
Four-channel: Auto, 3k pts, 30k pts, 300k pts, 3M pts and 6M
pts (option) are available
and 24M pts (option) are available
Dual-channel: Auto, 6k pts, 60k pts, 600k pts, 6M pts and
12M pts (option) are available
Four-channel: Auto, 3k pts, 30k pts, 300k pts, 3M pts and 6M
pts (option) are available
Input
Number of
Channels
Channels
four-channel
Input Coupling
DC, AC or GND
Input Impedance
(1 MΩ±2%) || (13 pF±3 pF)
Probe Attenuation
Coefficient
Coefficient
0.01X-1000X, 1-2-5 step
Max Input Voltage
(1MΩ)
(1MΩ)
Maximum input voltage of the analog channel
CAT I 300 Vrms, CAT II 100 Vrms, transient overvoltage
CAT I 300 Vrms, CAT II 100 Vrms, transient overvoltage