Acquisition Types 
Acquisition Types 
When you acquire a signal, the oscilloscope converts it into a digital form 
and displays a waveform. The acquisition sampling mode defines how the 
signal is digitized and the timebase setting affects the time span and level 
of detail in the acquisition. You can change the Sampling modes by 
pressing the ACQUIRE front panel button. 
There are three basic acquisition types: 
 
Sampling- samples the signal in evenly-spaced intervals to 
construct the waveform.  
Note
: You can use this type to reduce random noise. This type does 
not acquire rapid variation in the signal that may occur between 
samples. This can result in narrow pulses being missed. In this case, use 
Peak Detect
 to acquire data. 
 
Peak Detect - captures the maximum and minimum values that 
occur in a signal. It finds the highest and lowest record points over 
many acquisitions. 
Note:
 The oscilloscope can acquire and display narrow pulses, which 
may otherwise have been missed in Sampling mode. Noise appears 
higher in this mode. 
 
Average - acquires up to 256 waveforms, averages them, and 
displays the resulting waveform. 
Note
: This type can be used to reduce random noise. 
Use the following steps to choose a sampling type. 
1.  Press the ACQUIRE front panel button. On the Acquire menu, 
press the Acquisition option button and select a sampling type.