Lecroy WaveStation 2052 Function Generator, Frequency Generator WaveStation 2052 Manuale Utente

Codici prodotto
WaveStation 2052
Pagina di 32
WaveStation SCPI Command Reference Manual 
 
11 
 
920595 RevA 
BSWV 
– Basic Wave Command 
DESCRIPTION 
This command sets or retrieves basic waveform parameters.  
COMMAND
 
SYNTAX 
<channel>: BaSicWaVe <parameter> 
 
 
 
 
<channel>:= {C1,C2} 
  
 
 
 
<parameter>:= {A parameter from the following table} 
Parameters  Value 
Description 
WVTP 
<type> 
Sets the type of waveform. 
FRQ 
<frequency> 
Not applicable to Noise waveforms, this parameter sets a basic waveform’s 
frequency value. 
AMP 
<amplifier> 
Not applicable to Noise waveforms, this parameter sets a basic waveform’s 
amplification value. 
OFST 
<offset> 
Not applicable to Noise waveforms, this parameter sets a basic waveform’s 
offset value. 
SYM 
<symmetry> 
Only applying to Ramp waveforms, this parameter sets the waveform’s 
symmetry value.  
DUTY 
<duty> 
Only applying to Pulse waveforms, this parameter sets the waveform’s duty 
cycle percentage value. 
PHSE 
<phase> 
Not applicable to Noise waveforms, this parameter sets a basic waveform’s 
phase value. 
WAR 
<variance> 
Only applying to Noise waveforms, this parameter sets the waveform’s variance 
value. 
MEAN 
<mean> 
Only applying to Noise waveforms, this parameter sets the waveform’s mean 
value. 
DLY 
<delay> 
Only applying to Pulse waveforms, this parameter sets the waveform’s delay 
value. 
PLEASE NOTE THE FOLLOWING
 
If the command doesn’t set a basic waveform type, the parameters from the command are applied to the 
waveform type already selected on your WaveStation
.
 
 
Tables specifically covering command arguments are listed in the order in which they are functionally 
processed for convenience. 
Where: 
<type>:= {SINE,SQUARE,RAMP,PULSE,NOISE,ARB,DC} 
<frequency>:= {The default unit is in “HZ”. The minimum value is 1xe-6 HZ and 
the maximum depends on the WaveStation model.} 
<amplifier>:= {The default unit is in “V”. The C1 minimum value is 0.004V and 
maximum is 6V. The C2 minimum value is 0.004V and the maximum is 20V.} 
<offset>:= {The default unit is in “V”. The maximum value depends on your 
WaveStation model.} 
<duty>:= {If the waveform type is Square, the range is from 20 to 80. If the 
waveform type is Pulse, the range is from 10 to 90} 
<symmetry> := {0 to 100} 
<phase>:= {0 to 360}