Lecroy WaveStation 2052 Function Generator, Frequency Generator WaveStation 2052 Manuale Utente
Codici prodotto
WaveStation 2052
WaveStation SCPI Command Reference Manual
11
920595 RevA
BSWV
– Basic Wave Command
DESCRIPTION
This command sets or retrieves basic waveform parameters.
COMMAND
SYNTAX
<channel>: BaSicWaVe <parameter>
<channel>:= {C1,C2}
<parameter>:= {A parameter from the following table}
Parameters Value
Description
WVTP
<type>
Sets the type of waveform.
FRQ
<frequency>
Not applicable to Noise waveforms, this parameter sets a basic waveform’s
frequency value.
frequency value.
AMP
<amplifier>
Not applicable to Noise waveforms, this parameter sets a basic waveform’s
amplification value.
amplification value.
OFST
<offset>
Not applicable to Noise waveforms, this parameter sets a basic waveform’s
offset value.
offset value.
SYM
<symmetry>
Only applying to Ramp waveforms, this parameter sets the waveform’s
symmetry value.
symmetry value.
DUTY
<duty>
Only applying to Pulse waveforms, this parameter sets the waveform’s duty
cycle percentage value.
cycle percentage value.
PHSE
<phase>
Not applicable to Noise waveforms, this parameter sets a basic waveform’s
phase value.
phase value.
WAR
<variance>
Only applying to Noise waveforms, this parameter sets the waveform’s variance
value.
value.
MEAN
<mean>
Only applying to Noise waveforms, this parameter sets the waveform’s mean
value.
value.
DLY
<delay>
Only applying to Pulse waveforms, this parameter sets the waveform’s delay
value.
value.
PLEASE NOTE THE FOLLOWING:
If the command doesn’t set a basic waveform type, the parameters from the command are applied to the
waveform type already selected on your WaveStation
waveform type already selected on your WaveStation
.
Tables specifically covering command arguments are listed in the order in which they are functionally
processed for convenience.
processed for convenience.
Where:
<type>:= {SINE,SQUARE,RAMP,PULSE,NOISE,ARB,DC}
<frequency>:= {The default unit is in “HZ”. The minimum value is 1xe-6 HZ and
the maximum depends on the WaveStation model.}
the maximum depends on the WaveStation model.}
<amplifier>:= {The default unit is in “V”. The C1 minimum value is 0.004V and
maximum is 6V. The C2 minimum value is 0.004V and the maximum is 20V.}
maximum is 6V. The C2 minimum value is 0.004V and the maximum is 20V.}
<offset>:= {The default unit is in “V”. The maximum value depends on your
WaveStation model.}
WaveStation model.}
<duty>:= {If the waveform type is Square, the range is from 20 to 80. If the
waveform type is Pulse, the range is from 10 to 90}
waveform type is Pulse, the range is from 10 to 90}
<symmetry> := {0 to 100}
<phase>:= {0 to 360}