Lecroy WaveStation 2052 Function Generator, Frequency Generator WaveStation 2052 Manuale Utente
![Lecroy](https://files.manualsbrain.com/attachments/a6b425870750fec538437b1ab439f821a77f2154/common/fit/150/50/ff6974cbc973bf6214b742b12b227276bf1659fd6b79979bcdabfd971209/brand_logo.gif)
Codici prodotto
WaveStation 2052
WaveStation SCPI Command Reference Manual
21
920595 RevA
EXAMPLE
2
Set the modulation shape to AM, and set the AM type to a Sine wave.
C1: MDWV AM, MDSP, SINE
EXAMPLE
3
Read the C1 modulate waveform parameters confirming that STATE is ON.
C1: MDWV?
Return:
C1: MDWV
STATE,ON,AM,MDSP,SINE,SRC,INT,FRQ,100HZ,DEPTH,100,CARR,WVTP,RAMP,FR
Q,1000HZ,AMP,4V,OFST,0V,SYM,50
STATE,ON,AM,MDSP,SINE,SRC,INT,FRQ,100HZ,DEPTH,100,CARR,WVTP,RAMP,FR
Q,1000HZ,AMP,4V,OFST,0V,SYM,50
EXAMPLE
3
Read the C1 modulate waveform parameters confirming that STATE is OFF.
C1: MDWV?
Return:
C1: MDWV STATE,OFF
EXAMPLE
4
Set the C1 FM frequency to 1000HZ
C1: MDWV FM, FRQ, 1000HZ
EXAMPLE
5
Set the value of the C1 carrier waveform to SINE.
C1: MDWV CARR,MDSP,SINE
EXAMPLE
6
Set the value of the C1 carrier waveform frequency to 1000Hz.
C1: MDWV CARR,FRQ,1000HZ
RELATED
COMMANDS
ARWV, BTWV, CFG, CPL, SWWV, BSWV