Lecroy WAVEACE 2032 2-channel oscilloscope, Digital Storage oscilloscope, WaveAce 2032 Manuale Utente

Codici prodotto
WaveAce 2032
Pagina di 89
Operator's Manual 
920836 Rev A  
61 
STATUS - INR? 
Description 
The INR? query reads and clears the contents of the INternal state change Register (INR). The 
INR register records the completion of various internal operations and state transitions. 
Query Syntax 
INR? 
Response Format 
INR <value> 
<value>:= 0 to 65535 
Sample Response message: 
INR 1026 
Related Commands 
ALL_STATUS, *CLS 
INR Status Register Structure 
Bit 
Value 
Description 
15 
  
0 - Reserved for future use. 
14 
16384 
1 - Probe was changed. 
13 
8192 
1 - Trigger is ready. 
12 
4096 
1 - Pass/Fail test detected desired outcome. 
11 
2048 
1 - Waveform processing has terminated in trace F4. 
10 
1024 
1 - Waveform processing has terminated in trace F3. 
512 
1 - Waveform processing has terminated in trace F2. 
256 
1 - Waveform processing has terminated in trace F1. 
128 
1 - A disk exchange has been detected. 
64 
1 - Disk has become full in AutoStore Fill mode. 
32 
0 - Reserved for Teledyne LeCroy use. 
16 
1 - Segment of a sequence waveform has been acquired in acquisition memory but 
not yet read out into the main memory. 
1 - Time-out has occurred in a data block transfer. 
1 - Return to the local state is detected. 
1 - Screen dump has terminated. 
1 - New signal acquired in acquisition memory and read out to the main memory.