Single-shot Sampling Mode 
Basic Capture Technique 
A single-shot acquisition is a series of digitized voltage values sampled on 
the input signal at a uniform rate. It is also a series of measured data 
values associated with a single trigger event. The acquisition is typically 
stopped a defined number of samples after this event occurs: a number 
determined by the selected trigger delay and measured by the timebase. 
The waveform's horizontal position (and waveform display in general) is 
determined using the trigger event as the definition of time zero. 
You can choose either a pre- or post-trigger delay. Pre-trigger delay is the 
time from the left-hand edge of the display grid forward to the trigger 
event, while post-trigger delay is the time back to the event. You can 
sample the waveform in a range starting well before the trigger event up 
to the moment the event occurs. This is 100% pre-trigger, and it allows you 
to see the waveform leading up to the point at which the trigger condition 
was met and the trigger occurred. (The instrument offers up to the 
maximum record length of points of pre-trigger information.) Post-trigger 
delay, on the other hand, allows you to sample the waveform starting at 
the equivalent of 10,000 divisions after the event occurred. 
On fast timebase settings, the maximum single-shot sampling rate is used. 
But for slower timebases, the sampling rate is decreased and the number 
of data samples maintained. 
The relationship between sample rate, memory, and time can be simply 
defined as: 
Capture Interval = 1
/Sample Rate X Memory 
and 
Capture Interval
/10 = Time Per Division