Lecroy MSO64MXs-B 4-channel oscilloscope, Digital Storage oscilloscope, MSO64MXs-B Manuale Utente

Codici prodotto
MSO64MXs-B
Pagina di 128
G
ETTING 
S
TARTED 
M
ANUAL
 
49 
 WSMXs-B-GSM-E Rev A
 
TV 
 
TV triggers provide stable triggering on standard or custom composite 
video signals. Use them on PAL, SECAM, or NTSC systems. Support for 
HDTV triggering on 1080i, 1080p, and 720p formats is included.  Use 
CUSTOM setup for other standards.  
Serial
 
Serial trigger allows a serial trigger condition to be set from within the 
oscilloscope, using an easy-to-understand interface. 
Note: LeCroy offers a wide range of optional serial data triggering 
capabilities for serial data standards like I
2
C, SPI, UART, CAN, LIN, 
FlexRay, MIL-STD-1553, ARINC 429, MIPI D-PHY, DigRF, USB, and I
2
S. 
SMART Triggers 
 SMART Triggers are defined as follows: 
Glitch 
 
Glitch trigger is a simpler form of Width trigger. Use Glitch trigger for 
defining only fixed pulse-width times or time ranges. Glitch trigger 
makes no provision for voltage levels or ranges. 
Interval 
 
While Glitch trigger performs over the width of a pulse, Interval trigger 
performs over the width of an interval: the signal duration (the period) 
separating two consecutive edges of the same polarity (positive to 
positive or negative to negative). Use interval trigger to capture 
intervals falling short of or exceeding a given time limit. In addition, you 
can define a width range to capture any interval that is itself inside or 
outside the specified range, i.e. it can be used as an Exclusion trigger by 
interval. 
Dropout 
 
Used primarily in single-shot applications, and usually with a pre-trigger 
delay, Dropout trigger can detect lost signals. The trigger is generated at 
the end of the timeout period following the last trigger source 
transition. You can select a timeout period from 2 ns to 20 s, 
Runt 
 
The runt trigger is programmed to occur when a pulse crosses a first 
threshold line and fails to cross a second threshold line before 
recrossing the first. You can select both voltage thresholds within a time 
range of 100 ps to 20 s. Other defining conditions for this trigger are the 
edge (triggers on the slope opposite to that selected) and runt width.   
Slew Rate 
 
Slew rate trigger activates a trigger when the rising or falling edge of a 
pulse crosses two threshold levels: an upper level and a lower level. The 
pulse edge must cross the thresholds faster or slower than a selected 
period of time. You can select both thresholds within a range of 2 ns to 
20 s.