44
Trigger Setup
Trigger Delay
The trigger delay Indicator is a color-coded triangle
that is positioned horizontally on the grid.
Trigger delay horizontally positions the oscilloscope
trace on the display so you can observe the signal
prior to the trigger time. Delay adjusts the pre- and
post-trigger time. Pre-trigger delay is adjustable
from 0-100%
that is positioned horizontally on the grid.
Trigger delay horizontally positions the oscilloscope
trace on the display so you can observe the signal
prior to the trigger time. Delay adjusts the pre- and
post-trigger time. Pre-trigger delay is adjustable
from 0-100%
Post-trigger delay is indicated by a color-coded
arrow pointing to the left. Post-trigger delay is
adjustable from 0-10,000 divisions (calculated in
time)
arrow pointing to the left. Post-trigger delay is
adjustable from 0-10,000 divisions (calculated in
time)
All trigger delay values (including post-trigger
delay, shown here) are displayed in the Timebase
Descriptor Label.
delay, shown here) are displayed in the Timebase
Descriptor Label.
Zero delay is the horizontal center of the oscilloscope display.
Change trigger delay by using the D
Change trigger delay by using the D
ELAY
front panel knob in the HORIZONTAL
control area. Rotate the knob to adjust, or push to zero.
Trigger Level Indicator
The trigger level indicator is a color-coded triangle that is
positioned vertically on the grid.
In Stop trigger mode, the solid triangle also appears with
a hollow triangle of the same color if the level is changed
at that time. The hollow triangle indicates what the
trigger level is for the next valid trigger.
positioned vertically on the grid.
In Stop trigger mode, the solid triangle also appears with
a hollow triangle of the same color if the level is changed
at that time. The hollow triangle indicates what the
trigger level is for the next valid trigger.
Note: The Trigger Level indicator is not shown if the
triggered channel is not displayed.
triggered channel is not displayed.
Change trigger level using the front panel knob in the
Trigger Control area. Rotate the knob to adjust, or push
to set to 50% of the signal level.
Trigger Control area. Rotate the knob to adjust, or push
to set to 50% of the signal level.