Megger VDE-tester 1003-430 Scheda Tecnica

Codici prodotto
1003-430
Pagina di 5
An additional 250 V dc insulation test is available on all models 
for use on IT equipment, on devices or extension leads that are 
fitted with surge protection, or simply where it may be considered 
undesirable to use a 500 V test. 
Substitute leakage testing
Substitute leakage testing enables the measurements of AC leakage 
currents, which could differ significantly from the DC Insulation test 
results.
Substitute leakage testing is performed at less than 40 V ac. and 
does not require a mains supply.
Substitute leakage testing is used for measuring protective 
conductor current and touch current, in addition to an insulation 
test, or where it is considered an insulation test may damage 
equipment.
The Substitute leakage test will not operate equipment and so 
can be used where the operation of the equipment under test is 
undesirable.
Portable RCD testing (PRCD)
Portable RCDs can be tested using the PAT150. There is no need 
to find a non-RCD protected supply as the PAT150 will not trip 
external RCDs.
Both 10 mA and 30 mA RCDs can be tested both for disconnection 
times and the manual test button function.
Separated Extra Low Voltage (SELV) supplies
SELV supplies should be tested to ensure the output voltage 
does not exceed maximum limits as defined in the international 
regulations for Extra Low voltage systems. The PAT150 permits up 
to 50 V ac measurement, with a PASS or fail indication for SELV.
Differential Leakage (Protective Conductor Current), Touch 
Current & operational test
The PAT150 includes the facility for mains powered leakage test. 
This test has the benefit that equipment under test will function 
during the test sequence.
Battery life
The PAT120 & PAT150 operates from AA Alkaline or NiMH cells. 
Battery life is typically 4 days, based on 120 assets per day.
The PAT150R can be re-charged with NiMH batteries fitted.
2
speciFicAtions
enVironMentAL conDition:
operating ambient
Humidity
20°C
Nominal humidity
continUitY test
test voltage 
Compliance Voltage: +4V dc
–0 % /+10 % (open circuit)
test current
Bi-directional +200mA 
-0% + 50mA (into 2Ω load)
continuity accuracy
Resistance: ± 5% ±  3 digits (0 to 
19.99 Ω)
resistance resolution
10 mΩ
Display range
0.01 to 19.99Ω
continuity test nulling
up to 9.99 Ω
test time
User selectable from 2 sec to 20 
sec or selected during test to 
180sec
insULAtion test
insulation test
250V dc –0 % /+25 % open 
circuit
500V dc –0 % /+25 % open 
circuit
≥ 500V  0% dc across 0.5 MΩ 
load
short circuit/charge current < 2mA dc
insulation accuracy
±3% ±10 digits (0 to 19.99 MΩ)
resolution
0.01 MΩ
Display range
0.10 MΩ to 99.99 MΩ
test duration
User selectable from 2 sec to 20 
sec or selected during test to 
180sec
sUBstitUte LeAKAGe test
Leakage current
Accuracy ± 5% ± 3 digits
test frequency
Nominal mains frequency 50Hz
test voltage
< 50V ac
resolution
0.01mA
Display range
0.10 to 19.99mA
test duration
User selectable from 2 sec to 5 
seconds
reading corrected to 230V ac.
DiFFerentiAL LeAKAGe cUrrent
test voltage
Nominal supply voltage 230 V ac
test frequency
Nominal mains frequency 50  Hz
test accuracy
±5% ± 3 digits
resolution
0.01 mA
Display range
0.10 to 19.99 mA
test duration
User selectable from 2 sec to 5 
seconds