1507/1503 
Users Manual 
Table 3. Buttons and Indicators 
Button/ 
Indicator 
Description 
Press the blue button to select alternate 
measurement functions. 
Press to configure the Tester for a 
polarization index or dielectric absorption 
ratio test. The test will start when you press 
the 
T
 button. 
Sets a pass/fail limit for insulation tests.  
Test lock. When pressed before the 
T
 
button, the test remains active until you 
press the lock or test button again to 
release the lock.  
Turns the backlight on and off. The 
backlight goes off after 2 minutes. 
 
 
Button/ 
Indicator 
Description 
Initiates an insulation test when the rotary 
switch is an 
INSULATION
 position. Causes the 
Tester to source (output) a high voltage 
and measure insulation resistance. 
Initiates a resistance test when the rotary 
switch is in the ohms position. 
Unsafe voltage warning. Indicates 30 V or 
greater (ac or dc depending on the rotary 
switch position) is detected on the input. 
Also appears when the display shows 
0L
 in 
the 
b
 switch positions, and when 
batt
 
appears on the display. The 
Z
 also appears 
when insulation test is active. 
Pass indicator. Indicates when the 
insulation resistance measurement is 
greater than the selected compare limit.