Testboy TV 455VDE-tester Testboy TV 455 Manuale Utente

Codici prodotto
Testboy TV 455
Pagina di 24
TESTBOY TV 450 
Quick-test guide 
 
10 
2 Quick-test guide 
 
2.1  Null the leads 
 
Starting the tester: 
1.  Insert the fully charged batteries 
2.  Press the On/Off button 
3.  Connect test cable to the  
instrument. 
 
1
 Step 
 
‰
 
Select Continuity function (R LOWΩ or 
CONTINUITY*) with the function selector switch 
first. The following menu is displayed: 
 
 
 
 
2
 Step 
 
‰
 
Short test leads first as shown. 
 
N/L2
L/L1
L/L1
PE/L3
N/L2
PE/L3
prolongation lead
 
 
 
3
 Step 
 
‰
 
Press the TEST key in order to perform regular measurement. Result 
close to 0.00 Ω is displayed. 
 
 
4
 Step 
 
‰
 
Press the CAL key. After performing test leads compensation first measured 
value and then 0.00 Ω is displayed. If calibration was carried out successfully 
leads indicator 
CAL
 is shown in R LOWΩ and CONTINUITY* menus. 
 
                             
 
 
Measured value to be calibrated.                  
Reading at calibrated value is  
                                                                     0.00 Ω now. 
 
* The CONTINUITY subfunction is available in model TESTBOY TV 450 (Type B).