Chauvin Arnoux CA 6503 Insulation measuring device, P01132504 Manuale Utente

Codici prodotto
P01132504
Pagina di 57
English 
18 
C.A 6503 
3.2.  OPERATION: 
 
3.2.3. 
M
Ω
Ω
Ω
Ω
 measurement: 
 
M
Ω
Ω
Ω
Ω
 measurement at 250V
DC
The measurement is made at a voltage of 250V
DC
 above 1M
Ω
 measured. 
Short-
circuit current 
 5mA
DC
.
 
There is a change of reading coefficient for rising values: 50M
Ω
 (changes to X10 
reading), and a change of reading coefficient for falling values: 11M
Ω
 (changes to 
X1 reading). 
 
M
Ω
Ω
Ω
Ω
 measurement at 500V
DC 
The measurement is made at a voltage of 500V
DC
 above 1M
Ω
 measured. 
Short-
circuit current 
 5mA
DC
.
 
There is a change of reading coefficient for rising values: 50M
Ω
 (changes to X10 
reading), and a change of reading coefficient for falling values: 11M
Ω
 (changes to 
X1 reading). 
 
M
Ω
Ω
Ω
Ω
 measurement at 1000V
DC
The measurement is made at a voltage of 1000V
DC
 above 10M
Ω
 measured. 
Short-circuit current 
 5mA
DC
.
 
There is a change of reading coefficient for rising values: 500M
Ω
 (changes to X10 
reading), and a change of reading coefficient for falling values: 110M
Ω
 (changes 
to X1 reading).