C.A 6503
3.2. OPERATION:
3.2.3.
M
Ω
Ω
Ω
Ω
measurement:
M
Ω
Ω
Ω
Ω
measurement at 250V
DC
:
The measurement is made at a voltage of 250V
DC
above 1M
Ω
measured.
Short-
circuit current
≤
5mA
DC
.
There is a change of reading coefficient for rising values: 50M
Ω
(changes to X10
reading), and a change of reading coefficient for falling values: 11M
Ω
(changes to
X1 reading).
M
Ω
Ω
Ω
Ω
measurement at 500V
DC
:
The measurement is made at a voltage of 500V
DC
above 1M
Ω
measured.
Short-
circuit current
≤
5mA
DC
.
There is a change of reading coefficient for rising values: 50M
Ω
(changes to X10
reading), and a change of reading coefficient for falling values: 11M
Ω
(changes to
X1 reading).
M
Ω
Ω
Ω
Ω
measurement at 1000V
DC
:
The measurement is made at a voltage of 1000V
DC
above 10M
Ω
measured.
Short-circuit current
≤
5mA
DC
.
There is a change of reading coefficient for rising values: 500M
Ω
(changes to X10
reading), and a change of reading coefficient for falling values: 110M
Ω
(changes