For greatest accuracy, frequent calibration of the instrument is
recommended. The instrument should be recalibrated for pH:
a) Whenever the electrode is replaced.
b) At least once a month.
c) After testing aggressive chemicals.
d) Where extreme accuracy is required.
PREPARATION
Pour small quantities of pH 7.01 (
HI 7007) or pH 4.01 (HI 7004)
or pH 10.01 (
HI 7010) solution into a clean beaker.
To obtain accurate readings, use pH 7.01 (
HI 7007) if you are going
to measure neutral or close to neutral samples, pH 4.01 (
HI 7004)
if you are going to measure acidic samples or pH 10.01 (
HI 7010)
for alkaline measurements.
If you need to calibrate to NIST standards, use pH 6.86 (
HI 7006)
instead of pH 7.01 and pH 9.18 (
HI 7009) instead of pH 10.01.
PROCEDURE
• Connect the probe and switch the meter on, then press the 
pH
key to display pH measurement.
• Remove the protective cap from the probe, rinse and immerse it
in the buffer and stir gently. Wait a couple of minutes for the
reading to stabilize.
• Take the temperature of the buffer solution,
e.g. “
10.0°C”, as follow:
Select the °C mode and read the displayed
value.
• Adjust the pH calibration knob until the
LCD shows the pH value at the above
temperature (see the pH versus temperature
chart).
• The pH calibration is now complete.