Greisinger GTMU-MP-1 Digital Thermometer 605767 Manuale Utente

Codici prodotto
605767
Pagina di 6
T20.0.4X.6C-03 
 Operating Manual GTMU - MP 
page 6 of 6 
 
 
_____________________________________________________ 
_____________________________________________________________________________ 
 
11 Notes to the calibration services 
Calibration certificates - DKD-certificates - other certificates: 
If device should be certificated for its accuracy, it is the best solution to return it with the referring sensors to 
the manufacturer. (please state desired test values, e.g. 25°C) 
12 Specification 
Measuring range 
-
50.0 ... 400.0 °C  or  -58.0 ... 752.0 °F
 
Accuracy 
   Display 
  
Add. Output signal
 
(at nom. temperature 25°C) 
±0.4% of meas. value. ±0.2°C 
± 0.2% FS 
Media 
Non corrosive gasses 
Sensor 
Pt1000
 
Meas. frequency 
1 per second 
Adjusting 
Digital offset and scale adjustment for humidity and temperature 
Min-/Max-value memory 
Min and max measured values are stored 
Output signals 
refer to type plate,  
   Scaling 
both freely scalable by entering display values for 4mA (resp. 0V) and 20mA (resp. 
1V/10V) output 
   Connection 
  For option AV01, AV10 
4 - 20 mA    (2-wire) 
0 - 1 (10) V    (3-wire)  
  Auxiliary energy:  
   (supply voltage, each output) 
4 - 20 mA:   Uv = 12 - 30 V DC  
0-1V:  
Uv = 12 - 30 V DC, max. 10mA  
0-10V:  
Uv = 18 - 30 V DC, max. 10mA  
  Perm. impedance   
 
4 - 20 mA:   RA(Ohm) < ( (Uv - 12V) / 0,02A ) 
 
example: for Uv = 18V:  RA < (18V - 12V) / 0.02A     =>   RA < 300 Ohm 
  Permissible load 
 
0 - 1 (10) V: RL(Ohm) > 3000 Ohm 
 
Display 
approx. 10 mm high, 4-digit LCD-display 
Operating elements 
3 keys  
Ambient conditions  
   Nom. temperature 
   Operating temperature 
    
 
 
   Relative humidity 
   Storage temperature 
 
25°C 
Electronics: -25 ... 70 °C  
Please consider that the allowable range of the housing and electronics 
must not be exceeded, especially when using the probe at higher 
temperatures!
 
Electronics: 0 ... 95 %RH (not condensing) 
-25 ... 70 °C 
Housing 
ABS (IP65, except sensor head) 
   Dimensions 
82 x 80 x 55 mm (without elbow-type plug and sensor tube) 
for Option 
“Kabel”: Sensor head Ø14mm*68mm, 1m teflon cable, high humidity sensor 
   Mounting 
Holes for wall mounting (in housing - accessible after cover has been removed). 
   Mounting distance 
50 x 70 mm, max. shaft diameter of mounting screws is 4 mm 
   Electrical connection 
Elbow-type plug conforming to DIN 43650 (IP65), 
max. wire cross section: 1.5 mm², wire/cable diameter from 4.5 to 7 mm 
EMC 
The device corresponds to the essential protection ratings established in the 
Regulations of the Council for the Approximation of Legislation for the member 
countries regarding electromagnetic compatibility (2004/108/EG).  
In accordance with EN 61326-1
 
: 2006, additional errors: <1 % FS. When connecting 
long leads adequate measures against voltage surges have to be taken.