Keysight Technologies E5810B Scheda Tecnica

Pagina di 13
9
E5810B technical specifications 
General requirements
Minimum system requirements
Refer to page 4 for requirements in using the Agilent IO Libraries software (included 
with the instrument control hardware) 
Supported standards
GPIB Standard IEEE 488.1 and IEEE 488.2
USB 2.0 or lower (with the USBTMC-USB488 protocol)
LAN/Ethernet: 10BASE-T/100BASE-TX/1000BASE-T networks
RS-232
VXI-11 Protocol
VISA 2.2 and Agilent SICL
General characteristics
Power consumption
+12 VDC, 2 A
Isolated ELV supply source
Connectors
Standard 24-pin IEEE-488 (GPIB), USB 2.0, RS-232 (9-pin), LAN RJ-45
Maximum data rates
1.2 MB/s for GPIB
115 Kb/s for RS-232
480 Mb/s for USB
Maximum instrument connection
14 GPIB instruments 
1  USB  connection  (supports  up  to  4  USB  instruments  via  self-powered  hub)                                                                                                      
1 RS-232 instrument 
Up to 16 simultaneous connectivity connections
LED indicators
Power, LAN, GPIB, USB, RS-232, FAULT 
EMC and safety
IEC61326-1:2005 / EN61326-1:2006
Canada: ICES/NMB-001: Issue 4, June 2006
Australia / New Zealand: AS/NZS CISPR11:2004
IEC 61010-1:2010 / EN 61010-1:2010 (3rd Edition)
Warranty
3 years
Network protocols
See the E5810B User’s Manual for supported network protocols and functions
Dimensions
Width, depth and height
226.5 mm (8.92 in) × 238 mm (9.37 in) × 61mm (2.36 in)
Weight
1.3 kg
Environmental specifications
Operating environment
Operating temperature from 0 °C to 55 °C
Relative humidity up to 95% RH at 40 °C
Altitude up to 2000 m
Pollution Degree 2
Installation Category II (through an AC/DC adapter)
Storage environment
Storage temperature from –40 °C to 70 °C
Relative humidity up to 90% RH at 65 °C
Ordering information
Standard shipped accessories
Power cord and adapter                                                                                                                                          
Printed Quick Start Guide                                                                                
Agilent Automation-Ready CD with Agilent IO Libraries Suite 
Options
Option 100 – Rack Mount Kit 
GPIB cables/adapter (see page 11)