Gw Instek PCS-1000 Digital-Multimeter, DMM, PCS-1000 Scheda Tecnica
![Gw Instek](https://files.manualsbrain.com/attachments/dc6af9cceb6bec808125ac96ad99be7dbe37566d/common/fit/150/50/51850356515eec56d4ea258553f26e4c7c86af45115072a82f444014a577/brand_logo.gif)
Codici prodotto
PCS-1000
COMMUNICATION INTERFACE
69
Sense Commands
[SENSe:]CURRent:RANGe .......................................... 69
[SENSe:]CURRent:DC:AVERage:COUNt .................... 70
[SENSe:]CURRent:AC:AVERage:COUNt .................... 70
[SENSe:]VOLTage:RANGe .......................................... 70
[SENSe:]VOLTage:DC:AVERage:COUNt .................... 71
[SENSe:]VOLTage:AC:AVERage:COUNt .................... 72
[SENSe:]CURRent:RANGe
Set
Query
Description
Sets or queries the current range.
Syntax
Query Syntax
Query Syntax
[SENSe:]CURRent:RANGe {<Range>|AUTO}
[SENSe:]CURRent:RANGe?
[SENSe:]CURRent:RANGe?
Parameter /
Return Parameter
Return Parameter
<Range>
Current range <NRf>: 0.00000001~305
Sets the current range in amps. The unit
will automatically choose the closest
range that is programmed.
Sets the current range in amps. The unit
will automatically choose the closest
range that is programmed.
AUTO
Sets the range to AUTO; Only applicable
for the ≤3A ranges.
Autorange is not supported for the 30A
and 300A ranges.
for the ≤3A ranges.
Autorange is not supported for the 30A
and 300A ranges.
Example
CURR:RANG AUTO
Sets the current range to AUTO.
Sets the current range to AUTO.
Note
The range that is returned is the base unit. See the
table below:
table below:
Unit
Current Range
100
300A
10
30A
1
3A
.1
300mA
.01
30mA