Oxygen measuring 
The oxygen measurement depends on air pressure. The current pressure can be 
entered during configuration.  
If you measure at atmosphere and the calibration took place at the same 
conditions, the pressure value has not to be adjusted necessarily, because the 
same pressure is used for calibration and measuring.  
If the measurement is carried out at other pressure conditions than the 
calibration: The actual pressure has to be entered in the configuration for  
measurement and calibration.  
The sensor‟s temperature dependency is compensated internally. However for 
best results try to keep the same temperature both during calibration and 
measuring. 
Temperature differences between sensor and gas may cause additional errors! 
Therefore wait an appropriate period of time until the sensor has adjusted its 
temperature to the gas that is measured. A suitable flow of the gas around the 
sensor increases the adjustment significantly.  
Try to avoid warming of the sensor by touching it during measurement or 
calibration. 
When measuring bottled gas, consider that the gas coming out is cooler than the 
ambient because of the pressure loss! Strong air flows can produce an over 
pressure at the sensor 
– possible source of measuring/calibration errors! 
The nominal life time can be shortened by: 
Wrong storage / operation temperature 
Permanent use with dry gases (compressed gas). 
It helps to put the sensor to normal-humid ambient air in measuring breaks 
(“flush” system with fresh air). 
The optimum operation position is: with the sensor inlet pointing downwards 
Calibration 
In order to compensate for ageing of the sensor and air pressure fluctuations due 
to weather changes, the sensor has to be calibrated at regular intervals. We 
recommend to calibrate it at least once a week or, for optimum measuring results, 
directly before starting the measuring process. Adjust the absolute pressure at the 
configuration point “P.Ab” before carrying out a calibration. 
Calibration: The sensor will be calibrated to the atmospheric oxygen concentration 
of 20.95%
The sensor has to be subjected to air (make sure that rooms are 
thoroughly aired). 
Press the 
 key for 2 sec till 
 appears. 
The calibration will be automatically completed as soon as the measuring values 
for oxygen are stable (takes a few seconds).  After the calibration the device will 
show the rating of the sensor state for a short time. 
If an error message (CE.3, CE.4, CE.6) is displayed, the sensor signal is 
invalid. Press any key to restart, the previous calibration will be restored.