Operating Instructions
The U1586B temperature adapter is designed for Agilent handheld digital multimeter (DMM), Agilent handheld
oscilloscope, and K-Type thermocouple.
Assistance
For technical assistance, contact your nearest Agilent Sales Office or visit the Agilent website at
www.agilent.com/find/assist for further information.
Safety Informations
Please use the Agilent U1586B temperature adapter only as specified in this manual. Otherwise, the protection
provided may be impaired. Refer to the safety information below.
WARNING
To avoid possible electric shock, personal injury or damage to this instrument and ensure that you use the adapter
safely, follow these guidelines:
- Do not use the adapter if it is damaged. Before you use the adapter inspect the case. Look for cracks or missing
plastic. Pay particular attention to the insulation surrounding the connectors.
- Do not touch thermocouple probe to hazardous voltage.
- Do not operate the adapter around explosive gas, vapor or dust.
- When servicing the adapter, use only specified replacement parts.
- Removed all connections as battery replacement
- Use caution when working above 30 V ac rms, 42 V peak or 60 V dc. Such voltage pose a shock hazard.
- Avoid working alone.
- Do not operate the adapter as the cover removed or loosened.
CAUTION
To avoid possible damage to the temperature adapter or to the equipment under test, follow below guidelines:
- Do not connect output banana plugs to any power source.
- Use the proper terminals, function, and range for your measurements.
AC - Alternating Current
Caution, risk of danger (Refer to the user’s
and service guide for details)
Double Insulation
DC - Direct Current
Ground
Battery Voltage Test
Set the adapter’s slider switch to OFF position, the output will output the proportional mili-volt to represent the
battery voltage inside. The proportional scale is DC 9.9 mV per 1 V of battery voltage inside. The operating voltage
suggested is up to 7.5 V of battery. So, always make sure the mili-volt output is up to 74.2 mV before temperature
measurement.
Battery Voltage
10 V
9 V
8 V
7.5 V
Output of U1586B
99.0 mV
89.1 mV
79.2 mV
74.2 mV
Introductions
The U1586B temperature adapter accepts any K-Type thermocouple with miniature connector and converts it to 1
milli-volt per degree (Celsius or Fahrenheit).
Standard Item Purchased Checklist
The following items are included when you make a purchase:
•
U1586B temperature adapter
•
U1184A K-Type bead probe
•
Operation Instructions sheet (this sheet)
Conform to European Union Directives IEC/ EN
61010-12001 (2nd Edition)
OUTPUT: mV output proportional
to inside battery. °C and °F as the
position of slider switch
Slider switch:
• OFF: inside battery level
• °C: 1 mV/°C
• °F: 1 mV/°F
INPUT: K-type thermocouple with
miniature connector