Greisinger GLF 100RW Meter for reading performance ability GLF 100 RW ±(1% + 1 Count) -5 to +100 °C 600330 Scheda Tecnica
Codici prodotto
600330
H66.0.02.6C-03
Operating Manual GLF 100 RW
Page 3 of 8
1 Intended use
Precise measurements of conductivity in pure water applications with conductivities below
100µS/cm. Attention: Tap water is no pure water (several 100/µS/cm)!
Applications:
100µS/cm. Attention: Tap water is no pure water (several 100/µS/cm)!
Applications:
Monitoring of boiler water
Functional checking of ion exchangers
Checking of distilled / deionised / demineralised water
Functional checking of ion exchangers
Checking of distilled / deionised / demineralised water
etc.
2 General advice
Read through this document attentively and make yourself familiar to the operation of the device
before you use it. Keep this document in a ready-to-hand way in order to be able to look up in the
case of doubt.
before you use it. Keep this document in a ready-to-hand way in order to be able to look up in the
case of doubt.
3 Safety instructions
This device has been designed and tested in accordance to the safety regulations for electronic
devices. However, its trouble-free operation and reliability cannot be guaranteed unless the
standard safety measures and special safety advises given in this manual will be adhered to when
using it.
1. Trouble-free operation and reliability of the device can only be guaranteed if it is not subjected
devices. However, its trouble-free operation and reliability cannot be guaranteed unless the
standard safety measures and special safety advises given in this manual will be adhered to when
using it.
1. Trouble-free operation and reliability of the device can only be guaranteed if it is not subjected
to any other climatic conditions than those stated under “Specification”.
2. Transporting the device from a cold to a warm environment condensation may result in a failure
of the function. In such a case make sure the device temperature has adjusted to the ambient
temperature before trying a new start-up.
temperature before trying a new start-up.
3. The circuitry has to be designed most carefully if the device should be connected to other
devices. Internal connection in third party devices (e.g. connection GND and earth) may result
in not-permissible voltages impairing or destroying the device or another device connected.
in not-permissible voltages impairing or destroying the device or another device connected.
4. Whenever there may be a risk whatsoever involved in running it, the device has to be switched
off immediately and to be marked accordingly to avoid re-starting. Operator safety may be a
risk if:
- there is visible damage to the device or the device is not working as specified
- the device has been stored under unsuitable conditions for a longer time
In case of doubt, please return device to manufacturer for repair or maintenance.
risk if:
- there is visible damage to the device or the device is not working as specified
- the device has been stored under unsuitable conditions for a longer time
In case of doubt, please return device to manufacturer for repair or maintenance.
5.
Warning: Do not use this product as safety or emergency stop device or in any other
application where failure of the product could result in personal injury or material damage.
Failure to comply with these instructions could result in death or serious injury and material
damage
application where failure of the product could result in personal injury or material damage.
Failure to comply with these instructions could result in death or serious injury and material
damage
4 Operating and maintenance:
If the symbol "
BAT” is displayed at the left side of display, the battery is weak, measuring can
be continued for a short period. If
“bAt” is displayed in the main display the battery is used up
and needs to be replaced. Measuring is no more possible.
The battery has to be removed, when storing device above 50 °C.
Hint: We recommend removing the battery if device is not used for a longer period of time!
Risk of Leakage
Risk of Leakage
Treat device and probes carefully. Use only in accordance with above specification.
(do not throw, hit against etc.). Protect from soiling.
After switching on the instrument a segment test (all segments) is displayed,
followed by P.off, if auto power off function is activated (please refer to configuration).