Keysight Technologies Agilent U2802a Thermocouple Box 10 kS/s U2802A OPT ABD Scheda Tecnica
Codici prodotto
U2802A OPT ABD
60
[1]
Specifi cations are based on 30 minutes warm-up time and calibration temperature within 18 °C to 28 °C. In manual range, the settling time is 2.6
seconds while in autorange, the fi rst measurement accuracy is < 1%.
[2]
Specifi cations are for sine wave inputs more than 5% of range. 120% over range on all ranges except 250 VAC. Maximum crest factor of 5 at full
scale. Input impedance is 1 MΩ in parallel with capacitance less than 120 pF, AC couple with up to 300 VDC.
[3]
Additional error to be added as frequency more than 30 kHz and signal input less than 10% of range. 30 kHz to 100 kHz: add 0.003% of range per kHz.
[4]
Input signal has to be more than 50 Vrms.
[5]
Input protected with externally accessible 2 A, 250 V fast blown fuse.
AC specifi cations
[1]
AC accuracy for voltage
Function
Range
Accuracy input ±(% of reading + % of range)
Frequency (Hz)
20 ~ 45
45 ~ 10k
10k ~ 30k
30k ~ 100k
[3]
Voltage
[2]
100.000 mVrms
1 + 0.1
0.2 + 0.1
1.5 + 0.3
5.0 + 0.3
1.00000 V
1 + 0.1
0.2 + 0.1
1.0 + 0.1
3.0 + 0.2
10.0000 V
1 + 0.1
0.3 + 0.1
1.0 + 0.1
3.0 + 0.2
100.000 V
1 + 0.1
0.3 + 0.1
1.0 + 0.1
3.0 + 0.2
250.000 V
[4]
1 + 0.1
0.3 + 0.1
1.0 + 0.1
3.0 + 0.2
Temperature coeffi cient for voltage (0 ºC to 18 ºC, 28 ºC to 55 ºC)
Function
Range
Frequency (Hz)
(% of reading + % of range)
20 ~ 45
45 ~ 10k
10k ~ 30k
30k ~ 100k
Temperature
coeffi cient
100.000 mVrms
0.02 + 0.02
0.02 + 0.02
0.05 + 0.02
0.1 + 0.02
1.00000 V
10.0000 V
100.000 V
250.000 V
AC accuracy for current
[1]
Function
Range
Burden voltage,
Current shunt
resistance
Accuracy input ±(% of reading + % of range)
Frequency (Hz)
20 ~ 45
45 ~ 1k
1k ~ 10k
Current
[5]
10.0000 mA
< 0.2 V, 10 Ω
1.5 + 0.1
0.5 + 0.1
2 + 0.2
100.000 mA
< 0.2 V, 1 Ω
1.5 + 0.1
0.5 + 0.1
2 + 0.2
1.00000 A
< 0.3 V, 0.1 Ω
1.5 + 0.1
0.5 + 0.1
2 + 0.2
2.00000 A
< 0.8 V, 0.1 Ω
1.5 + 0.1
0.5 + 0.1
2 + 0.2
Temperature coeffi cient for voltage (0 ºC to 18 ºC, 28 ºC to 55 ºC)
Function
Range
Accuracy input ±(% of reading + % of range)
Frequency (Hz)
20 ~ 45
45 ~ 10k
10k ~ 30k
Temperature
coeffi cient
10.0000 mA
0.02 + 0.02
0.02 + 0.02
0.02 + 0.02
100.000 mA
1.00000 A
2.00000 A
NOTE
To ensure better measurement results and to guard against the change of environment or setup, always
enable the Null offset.
enable the Null offset.