Keysight Technologies USB data capturing module U2354A Input: 500 kS/s U2354A OPT 902 + ABD Scheda Tecnica

Codici prodotto
U2354A OPT 902 + ABD
Pagina di 75
23
Accuracy Specifi cations (Continued)
Model
U2331A
Bipolar Range (V)
Offset Error (mV)
[1]
Gain Error (mV)
Accuracy (% of reading + offset error)
[2]
10
2.0
6.0
0.06% + 2.0 mV
5
1.5
4.0
0.08% + 1.5 mV
2.5
1.5
2.0
0.08% + 1.5 mV
1.25
1.0
1.5
0.12% + 1.0 mV
1
1.0
1.0
0.10% + 1.0 mV
0.5
1.0
1.0
0.20% + 1.0 mV
0.25
1.0
1.0
0.40% + 1.0 mV
0.2
1.0
1.0
0.50% + 1.0 mV
0.05
1.0
1.0
2.02% + 1.0 mV
•  The above specifi cations are typical for 23 °C.
• Specifi cations are for 20 minutes warm-up and self calibration.
•  The measurement are calculated with 100 points averaging of data.
Model
U2351A, U2352A, U2353A, U2354A
Analog output
Unipolar Range (V)
Offset Error (mV)
[3]
Gain Error (mV)
Accuracy (% of reading + offset error)
[4]
10
1.0
2.0
0.02% + 1.0 mV
Bipolar Range (V)
Offset Error (mV)
[3]
Gain Error (mV)
Accuracy (% of reading + offset error)
[4]
10
1.0
4.0
0.04% + 1.0 mV
Model
U2355A, U2356A
Unipolar Range (V)
Offset Error (mV)
[3]
Gain Error (mV)
Accuracy (% of reading + offset error)
[4]
10
1.0
2.0
0.02% + 1.0 mV
Bipolar Range (V)
Offset Error (mV)
[3]
Gain Error (mV)
Accuracy (% of reading + offset error)
[4]
10
1.0
4.0
0.04% + 1.0 mV
Model
U2331A
Unipolar Range (V)
Offset Error (mV)
[3]
Gain Error (mV)
Accuracy (% of reading + offset error)
[4]
10
2.5
4.0
0.04% + 2.5 mV
Bipolar Range (V)
Offset Error (mV)
[3]
Gain Error (mV)
Accuracy (% of reading + offset error)
[4]
10
1.5
4.0
0.04% + 1.5 mV
•  The above specifi cations are typical for 23 °C.
• Specifi cations are for 20 minutes warm-up and self calibration.
[1]
 Offset error is measured at midscale of full range.
[2]
 Accuracy = ± [% of |(Gain error / (Measured value – Midscale of FSR))| + Offset error]
[3]
 Offset error is measured at 0 V.
[4]
 Accuracy = ± [% of |Gain error/Output value| + Offset voltage]