Keysight Technologies USB data capturing module U2354A Input: 500 kS/s U2354A OPT 902 + ABD Scheda Tecnica

Codici prodotto
U2354A OPT 902 + ABD
Pagina di 75
59
Product specifi cations
DC specifi cations
[1]
Function
Range
Input impedance
Test current/
Burden voltage, 
Shunt resistance
Accuracy 
±(% of reading + 
% of range)
Temperature 
coeffi cient 
0 °C to 18 ºC
28 °C to 55 °C
Voltage
[2]
100.000 mV
10 MΩ
-
0.015 + 0.008
0.002 + 0.0008
1.00000 V
10 MΩ
-
0.015 + 0.005
0.001 + 0.0005
10.0000 V
10 MΩ
-
0.018 + 0.005
0.002 + 0.0005
100.000 V
10 MΩ
-
0.018 + 0.005
0.002 + 0.0005
300.000 V
10 MΩ
-
0.018 + 0.005
0.002 + 0.0005
Current
[3]
10.0000 mA
-
< 0.2 V, 10 Ω
0.06 + 0.015
0.005 + 0.0025
100.000 mA
-
< 0.2 V, 1 Ω
0.06 + 0.005
0.008 + 0.002
1.0000 A
-
< 0.3 V, 0.1 Ω
0.15 + 0.007
0.005 + 0.002
2.0000 A
-
< 0.8 V, 0.1 Ω
0.15 + 0.007
0.005 + 0.002
Resistance
[4]
100.000 Ω
-
1.0 mA
0.03 + 0.008
0.006 + 0.0008
1.00000 kΩ
-
1.0 mA
0.03 + 0.005
0.006 + 0.0005
10.0000 kΩ
-
100 μA
0.03 + 0.005
0.006 + 0.0005
100.000 kΩ
-
10.0 μA
0.03 + 0.005
0.006 + 0.0005
1.00000 MΩ
-
1 μA
0.06 + 0.005
0.01 + 0.0005
10.0000 MΩ
-
225 nA
0.25 + 0.005
0.025 +0.0005
100.000 MΩ
-
225 nA || 10 MΩ
2.0 + 0.005
0.3 + 0.0005
Diode test
[5]
1.0000 V
-
1.00 mA
0.015 + 0.03
0.005 + 0.0005
Continuity test
[6]
1.0000 kΩ
-
1.00 mA
0.05 + 0.03
0.005 + 0.0005
NOTE
To ensure better measurement results and to guard against the change of environment or setup, always 
enable the Null offset.
[1]
 Specifi cations are based on 30 minutes warm-up time, NPLC 20 resolution, and calibration temperature within 18 °C to 28 °C. For NPLC 0 and 0.025, 
add 0.01% of range.
[2]
 120% over range on all ranges except 300 VDC. Input protection up to 300 VDC.
[3]
 Input protected with externally accessible 2 A, 250 V fast blown fuse.
[4]
 Specifi cations are for 4-wire Ω or 2-wire Ωs using null function in AMM software. If without null function in AMM software, add 0.2 Ω additional error. 
Input protection up to 300 VDC. Specifi cations apply for NPLC ≥ 1.
[5]
 Specifi cations are for the voltage measured at the input terminals only.
[6]
 Continuity threshold is fi xed at less than 10 Ω.