Keysight Technologies USB data capturing module 2541A Input: 250 kS/s, output: 1 MS/s U2541A OPT 902 + ABD Scheda Tecnica

Codici prodotto
U2541A OPT 902 + ABD
Pagina di 75
42
Table 3.  Measurement accuracy of the U2356A at 0 to 18 °C and 28 to 45 °C
Thermocouple measurement accuracy (U2356A @ 0 to 18 °C and  28 to 45 °C)
T/C type
ITS-90 Temperature range (°C)
Optimum measurement range 
(°C)
Without 
averaging
(± °C)
50 points 
averaging
(± °C)
500 points 
averaging
(± °C)
Low
High
Low
High
B
0
1820
1100
1820
6.1
3.1
2.4
400
1100
14.4
6.3
2.7
E
–270
1000
–150
1000
3.0
2.6
2.6
–200
–150
4.2
3.7
3.6
J
–210
1200
–150
1200
2.9
2.5
2.5
–210
–150
4.9
4.1
4.0
K
–270
1372
–100
1200
3.3
2.9
2.9
–200
–100
5.3
4.2
4.0
N
–270
1300
–100
1300
3.4
2.7
2.6
–200
–100
6.8
4.6
4.1
R
–50
1768
300
1760
6.2
3.7
3.2
–50
300
15.7
7.2
3.8
S
–50
1768
400
1760
6.4
4.0
3.4
–50
400
14.2
6.6
3.4
T
–270
400
–100
400
3.0
2.4
2.2
–200
–100
5.3
4.2
3.9