Rigol DSA800-AMK Scheda Tecnica
RIGOL
DSA800 Options and Accessories
12
1.5 GHz Tracking Generator
DSA800-TG option provides tracking output function. When the tracking generator is
enabled, a signal with the same frequency of the current sweep signal is output from
the [GEN OUTPUT 50 Ω] connector at the front panel and the power of the signal
can be set through the menu. The tracking generator provides two working modes:
Power Sweep output and Fixed Power output.
When the tracking generator is used as an excitation, the DUT (device under test)
characteristics can be obtained by measuring the output from the DUT. For example,
use the tracking generator to verify the frequency response of the filter. Connect the
filter to be tested to the [RF INPUT 50Ω] and [GEN OUTPUT 50Ω] connectors at
the front panel of the spectrum analyzer.
Filter