Electrical Specifications (Ambient Temperature 23°C ± 5°C) 
Air velocity 
Measurement 
Range 
Resolution 
Accuracy 
m/S 
0.4 - 25.0 m/s 
0.1 m/S 
± (2% + 0.2 m/S) 
Km/h 
1.4 - 90.0 km/h 
0.1 Km/h 
± (2% + 0.8 Km/h) 
mph (Mile/h) 
0.9 - 55.9 mile/h 
0.1 mph 
± (2% + 0.4 mph) 
Knots 
0.8 - 48.6 knots 
0.1 Knots 
± (2% + 0.4 Knots) 
FPM (Ft/min) 
80 - 4930 ft/min 
1 FPM 
± (2% + 40 FPM) 
Air temperature 
Measuring Range 
0 to 50°C (32 to 122°F) 
Resolution 
0.1°C (0.1 °F) 
Accuracy 
± 0.8°C (1.5°F) 
 
Type K/J Thermocouple Thermometer Temperature 
 
Sensor Type Resolution 
Range 
Accuracy 
Type K 
0.1°C 
-50.0 to 1300.0°C 
-100.0° to -50.1C 
± (0.4%+ 0.5°C) 
± (0.4%+ 1°C) 
0.1°F 
-58.0 to 2372.0°F 
-148.0 to -58.1°F 
± (0.4%+ 1°F) 
± (0.4%+ 1.8°F) 
Type J 
0.1°C 
-50.0 to 1200.0°C 
-100.0 to -50.1°C 
± (0.4%+ 0.5°C) 
± (0.4%+ 1°C) 
0.1°F 
-58.0 to 2192.0°F 
-148.0 to -58.1°F 
± (0.4%+ 1°F) 
± (0.4%+ 1.8°F) 
 
Note: Above specifications tested under an environmental RF Field Strength lower than 3 V/M and a 
frequency lower than 30 MHz  
 
 
 
 
 
 
 
 
 
 