Calibration
2
U2300A Series DAQ Service Guide
21
The following figure illustrates the connection of a floating 
signal source to the U2300A series DAQ in differential input 
mode. For floating signal sources, additional resistor is 
needed at each channel to provide a bias return path. The 
resistor value is equivalent to about 100 times the source 
impedance. If the source impedance is less than 100 W, you 
can connect the negative polarity of the signal directly to 
AI_GND, as well as the negative input of the Instrumentation 
Amplifier. The noise couples in differential input mode are 
less compared to the single- ended mode.
Figure 2-4
Floating source and differential input
N O T E
• Agilent U2300A series DAQ is designed with high input impedance. 
Please ensure that all the connection are connected properly before 
acquiring any data. Failing to do so may cause data fluctuation or 
erroneous readings.
• Unused pins at multiplexing DAQ inputs can be treated as floating 
source with infinite output impedance. Therefore, necessary grounding 