Troubleshooting and Self-Test Procedures
U2300A Series DAQ Service Guide
49
Self-Test Procedures
4 Go to Start > All Programs > Agilent IO Libraries Suite > Agilent
Connection Expert to launch the Agilent Connection Expert.
5 Go to Start > All Programs > Agilent T&M Toolkit > Agilent Interactive IO
to launch the Interactive I/O dialog box.
6 Send the SCPI command “
*TST?” to the instrument to start
perform the self- test of the instrument.
7 The command will return either "
+0" to indicate all tests passes
or "
+1" to indicate one or more tests failed.
8 If the command returns "
+1" , apply SCPI command
“
SYSTem:ERRor?” to enquire the error message.
WA R N I N G
Do not connect any cables and terminal block prior to performing
self-test procedures.
self-test procedures.
N O T E
Refer to Agilent U2300A Series USB Multifunction Data Acquisition
Programming Guide for SCPI error messages.
Programming Guide for SCPI error messages.