Troubleshooting and Self-Test Procedures
U2300A Series DAQ Service Guide
49
Self-Test Procedures
Go to Start > All Programs > Agilent IO Libraries Suite > Agilent 
Connection Expert to launch the Agilent Connection Expert.
Go to Start > All Programs > Agilent T&M Toolkit > Agilent Interactive IO 
to launch the Interactive I/O dialog box.
Send the SCPI command “
*TST?” to the instrument to start 
perform the self- test of the instrument.
The command will return either "
+0" to indicate all tests passes 
or "
+1" to indicate one or more tests failed.
If the command returns "
+1" , apply SCPI command 
SYSTem:ERRor?” to enquire the error message.
WA R N I N G
Do not connect any cables and terminal block prior to performing 
self-test procedures.
N O T E
Refer to Agilent U2300A Series USB Multifunction Data Acquisition 
Programming Guide
 for SCPI error messages.