Helios Preisser DIGI-MET® Instrument 1726 502 Reading range(s) 12.5 mm Reading 0,001 mm Accuracy 0,005 mm 1726 502 Scheda Tecnica
Codici prodotto
1726 502
15
UsbCom Standard / Professional 2.1
©
Software
With the keyboard code settings you can define 2 end markers and there number. Once a measured value
has been transferred the sequence of the End markers will be also be sent. For example, if ENTER has been
selected, this corresponds to pressing the enter key on the keyboard (measured value is transferred and the
position of the measured value depends upon the chosen End markers / sequence). In this way you can jump
to any field in your form or table after a measured value has been transmitted.
has been transferred the sequence of the End markers will be also be sent. For example, if ENTER has been
selected, this corresponds to pressing the enter key on the keyboard (measured value is transferred and the
position of the measured value depends upon the chosen End markers / sequence). In this way you can jump
to any field in your form or table after a measured value has been transmitted.
In order to select the End marker click on the ENTER field and select the required character from the Pull-
Down menu.
Down menu.
With the decimal separator select the decimal separator according the type used in the country you are in or
according to type used in the application.
according to type used in the application.
With „keyboard code“ transfer to, the measured values of the measuring instrument are transferred into the
actual current position of the cursor. Therefore, it is possible to transmit the measured value independently of
the software application. Measured values can therefore be transferred into MS Access, OpenOffice or into a
CAQ program.
actual current position of the cursor. Therefore, it is possible to transmit the measured value independently of
the software application. Measured values can therefore be transferred into MS Access, OpenOffice or into a
CAQ program.