The operation state register's bit meanings are:
Bit
Meaning
0
Calculate the new demarcation coefficient
1
Waiting for a trigger signal
2
Remote control state (1 means enabled)
3
Output state (1 means ON)
4
Local key state (0 means not enabled, 1 means enabled)
5
Remote sensing mode (1 means enabled)
6
LOAD ON timer is enabled
7
Reserved
The demand state register's bit meanings are:
Bit
Meaning
0
Reversed voltage is at instrument's terminals (1 means yes)
1
Over voltage (1 means yes)
2
Over current (1 means yes)
3
Over power (1 means yes)
4
Over temperature (1 means yes)
5
Not connect remote terminal
6
Constant current
7
Constant voltage
8
Constant power
9
Constant resistance
0x60 Enter instrument calibration state
Byte
offset
Meaning
3
Calibration state:
0 means disable
1 means enable
1 means enable
4
0x85 (low byte of calibration password)
5
0x11 or 0x12 (high byte of calibration password)
6-24
Reserved
0x61 Get the instrument calibration state
8500 DC Load Series
Version: September 2, 2009
Page 67 of 76