EN
Instruction Manual
EL 9000 HP Series
Date: 10-31-2011
6.2  Operating elements
Mains switch 
Power (1)
Is used to switch the device on or off. 
Selector 
Mode (2)
Is used to preselect the regulation 
mode  in  which  the  device  shall 
work.  The  different  regulation 
modes effect each other. As soon 
as actual values reach set values, 
the regulation mode changes. This 
can, for example, lead to a dominance of the constant 
power control (CP), even if the constant current control 
(CC) is set as active mode. The currently active regu-
lation mode is indicated in the display. More about the 
different regulation modes in section „6.6. Preselecting 
the regulation mode“.
Following regulation modes are selectable:
CC     
Constant current regulation
CV     
Constant voltage regulation
CP     
Constant power regulation
CR
10R
  
Constant resistance regulation in range 1*
CR
400R
 
Constant resistance regulation in range 2*
*the shown range values are only examples
The switchover of the regulation mode switches the load 
input off and the set value that is related to the currently 
chosen  regulation  mode  is  addressed  for  adjustment 
(arrow in front of the value). Other set values can also be 
accessed and adjusted in the described manner. 
Note: Preselection of CV enable adjustment of the voltage 
set value. Same applies for the resistance value and CR.
Selector 
Level Control (3)
Is used to select the operation mode  
between  normal  operation  mode 
(
A and B), pulsed mode (A/B) and 
battery test mode. All 4 modes have 
an own set of set values. That means, 
that you can adjust different set va-
lues for U, I, P etc. for 
A than for B. When selecting a 
different regulation mode with 
Mode(2) the set values 
will be reset to certain default values, if the setting 
Keep
set values
 is set to 
no
details). Using 
Level Control(3) switches the load input 
off, except if changed from 
A->B or B->A, because these 
positions are used to manually switch between the two 
levels.
The  most  recent  set  values  are  stored  inside  the  unit 
when the unit is switched off or at mains undervoltage 
(power fail), and are restored after switching it on again, 
if the setting 
Keep set values = yes
 has been chosen.
Explanation of the selector positions:
A      Switches to the set values of Level A. These 
values  become  instantly  active  and  can  be 
changed now. 
B      Switches to the set values of Level B. These 
values  become  instantly  active  and  can  be 
changed now.
A/B    Switches  the  load  offline  and  activates  the 
pulsed mode (frequency controlled, automatic 
changeover between level A and B).
Battery  Switches  the  load  offline  and  activates  the 
battery test mode. 
Setup    Switches  the  load  offline  and  activates  the 
setup menu. Here the device and the interface 
card (if equipped) are configured.
More about the operation modes in section „6.7. Usage 
Pushbutton
Input on/off (4)
Is used to activate or deactivate load operation 
(i.e. input on/off). Input off is indicated in the dis-
play with 
Standby
. Switching the input on can 
be inhibited by several reasons. For example, 
if an error is persistent or the input is blocked 
by the REM-SB pin of the analogue.
In battery test mode, the time counting is halted or conti-
nued if the button is pressed to switch the input on or off.
It is also used to acknowledge the alarm indication in 
the display. By pushing the button, the indicated error is 
cleared (as long as no alarm is still active) and the button 
can be used as normal.
In  remote  operation,  i.e.  control  via  a  digital  interface 
card, the button can be used to forcibly reset from remote 
operation by pressing >3s.
New since Firmware 5.01: in mode „Battery“, the button 
can be used to reset time and Ah counter, if pressed >3s 
in offline state.
Rotary knob 
Selection (5)
This rotary knob has no end stop. With every 
position  
Selection(5) selects another ele-
ment in the display resp. another parameter 
in the setup menu. An arrow (->)is placed 
in front of the selected element/parameter.
The selection sequence is, when rotating the knob clo-
ckwise, top left -> bottom left -> top right -> bottom right. 
That means, if the arrow is pointing to the voltage set value 
it will be moved to the set value for current and then to the 
power set value etc. 
Counter-clockwise rotation moves the arrow in reversed 
order.
Handling the device